CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layer 1 Replacement Elog  Not logged in ELOG logo
Message ID: 152     Entry time: Fri Mar 27 17:27:09 2020
Author: Andrey Starodumov 
Category: Reception test 
Subject: M1635 failed Reception 
All ROCs are programmable but DESER400 trailer error bits: "NO DATA" or "IDLE DATA".
ROC8-11 are affected (no data)
Visual inspection is OK
To module doctor
ELOG V3.1.3-7933898