Message ID: 278     Entry time: Fri May 29 13:56:35 2020
Author: Andrey Starodumov 
Category: Re-grading 
Subject: M1630 
In ROC1 of M1630 gain calibration failed massively (3883 pixels) at +10C with CtrlReg 9.
A special test of M1630 only at +10C and with CtrlReg 17 showed no problem.
p10_1 from ~/L1_DATA/ExtraTests_ToBeKept/M1630_FullTestp10_2020-04-17_15h54m_1587131688/000_Fulltest_p10
copied to ~/L1_DATA/M1630_FullQualification_2020-04-06_08h35m_1586154934/005_Fulltest_p10
and original files from ~/L1_DATA/M1630_FullQualification_2020-04-06_08h35m_1586154934/005_Fulltest_p10
copied to ~/L1_DATA/ExtraTests_ToBeKept/p10RemovedFrom_M1630_FullQualification_2020-04-06_08h35m_1586154934/005_Fulltest_p10
This is done to have a clean ranking of modules based on # of defects.