Message ID: 278
    Entry time: Fri May 29 13:56:35 2020
 |  
| Author: | 
Andrey Starodumov  | 
 | Category: | 
Re-grading  | 
 | Subject: | 
M1630  | 
 
  |  
 
 | 
In ROC1 of M1630 gain calibration failed massively (3883 pixels) at +10C with CtrlReg 9. 
A special test of M1630 only at +10C and with CtrlReg 17 showed no problem. 
p10_1 from ~/L1_DATA/ExtraTests_ToBeKept/M1630_FullTestp10_2020-04-17_15h54m_1587131688/000_Fulltest_p10 
copied to ~/L1_DATA/M1630_FullQualification_2020-04-06_08h35m_1586154934/005_Fulltest_p10 
and original files from  ~/L1_DATA/M1630_FullQualification_2020-04-06_08h35m_1586154934/005_Fulltest_p10 
copied to ~/L1_DATA/ExtraTests_ToBeKept/p10RemovedFrom_M1630_FullQualification_2020-04-06_08h35m_1586154934/005_Fulltest_p10 
This is done to have a clean ranking of modules based on # of defects. |