Message ID: 279     Entry time: Fri May 29 15:04:04 2020
Author: Andrey Starodumov 
Category: XRay HR tests 
Subject: Analysis of HRT: M1555-M1561 and M1564 
Module HRtest VCal calibration Grade
#defects max #noisy pix
ColdBox XRay
M1555: 24 44 375 44xVcal-388e- A
M1556: 54 74 507 43xVcal-370e- B
M1557: 109 63 145 43xVcal-216e- A
M1558: 127 113 103 44xVcal-145e- B
M1559: 69 59 92 46xVcal-119e- A
M1560: 35 54 66 46xVcal-123e- A
M1561: 33 44 129 45xVcal-244e- A
M1564: 26 36 93 47xVcal- 29e- A