CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layers 2-4 Elog  Not logged in ELOG logo
Message ID: 4     Entry time: Tue Sep 8 10:28:08 2020
Author: Matej Roguljic 
Category: Qualification 
Subject: Qualification on 07.09. 
Four L2 modules were tested. The tests consist of Fulltest@-20, IV@-20, Temperature cycle (n=3), Fulltest@+10, IV@+10.
Modules tested:
M2026: A
M2078: A
M2126: A
M2160: C - this one is interesting since it's graded A at +10, but at -20 the results are horrible and we can see sensor wafer pattern.