CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layer 1 Replacement Elog  Not logged in ELOG logo
Message ID: 129     Entry time: Tue Mar 24 15:41:29 2020
Author: Andrey Starodumov 
Category: Module grading 
Subject: Change in MoreWeb GradingParameters.cfg  
On March 3d the Vcal calibration parameter has been changed:
- StandardVcal2ElectronConversionFactor from 50 to 44 (electrons/VCal)

Om March 24 the following changes done:
- trimThr from 35 to 50 (to synchronize with current target trimming threshold of Vcal=50)
- TrimBitDifference from 2. to -2. This means that difference between trimmed and untrimmed
threshold close to 0 (<2 as it was) will be ignored.