CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layer 1 Replacement Elog  Not logged in ELOG logo
Message ID: 58     Entry time: Mon Feb 3 11:28:38 2020
Author: Anrey Starodumov 
Category: Full test 
Subject: 4 modules FT: 1536, 1537,1538, 1540 
Jan 28: Reception test for 1536, 1537, 1538 OK
Then cap gluing
Jan 29
- Reception test for 1540 OK
- FT:
-- 1536: B/B/B (-20/-20/+10). Electrical Test is B due to mean noise >200e. IV: 3umA for -20 and +10, slope >300
-- 1537: C/B/B. First -20 is C due ROC10 >200 TrimBits failure, noise >200e. IV: A
-- Trimbit test fails but Trimbit distribution is OK and Threshold reasonable (attachment)
==> Manually regraded to B
-- 1538: B/B/B (-20/-20/+10). Electrical Test is B due to mean noise >200e. IV: <0.2umA for +10, slope ~1
-- 1540: B/B/B (-20/-20/+10). Electrical Test is B due to mean noise >200e. IV: <0.8umA for +10, slope ~2.7
Attachment 1: 1537ROC10m20-TrThr.pdf  14 kB  Uploaded Mon Feb 3 15:05:30 2020  | Hide | Hide all
1537ROC10m20-TrThr.pdf
ELOG V3.1.3-7933898