Entry  Mon May 11 14:40:16 2020, danek kotlinski, Other, M1582 m1582_roc1_thr_1d.pngm1582_roc1_thr_2d.pngm1582_roc1_ph70.png
On Friday I have tested the module M1582 at room temperature in the blue box.
The report in MoreWeb says that this module has problems with trimming 190 pixels in ROC1.
Entry  Thu Feb 27 10:14:09 2020, danek kotlinski, Module transfer, M1562 to DESY 
M1562 was not qualified during a full test since the test did not complete.
Looking at the module I see that roc 2 has a lot of noise row 79.
I tried to mask all pixels in this row but the masking does not work.
Entry  Fri Aug 28 11:53:23 2020, Andrey Starodumov, PhQualification, M1555 
There is no new PH optimisation for this module?!
To be checked!
Entry  Fri May 29 15:41:40 2020, Andrey Starodumov, General, M1546 C10-p10.pdfTrimBits.pdfC10-50.pdf
ROC10 of M1546 has 107 pixels with trimming problems:
in a VCal threshold scan after trimming 81 has a threshold underflow, means <0, and 26 pixels are outside 40-60VCal window (around Vcal 50) at +10C only.
Trimbit distribution looks reasonable.
Entry  Tue Mar 31 06:54:12 2020, danek kotlinski, Module grading, M1542 Canvas_1.png
I have tested M1542.
It looks not bad but there is a group of pixels in the lower/left corner of ROC6 which behave different.
One can also see that something is not right in some PH optimisation plots.
Entry  Wed Apr 29 08:48:06 2020, Urs Langenegger, Other, M1539 
M1539 showed no readout. I tried, all without success, 
- reconnecting the cable to the adapter multiple times
- connecting to the adapter in the blue box
Entry  Mon May 11 13:19:51 2020, Andrey Starodumov, Cold box tests, M1539 
After several attempts including reconnecting the cable, M1539 had no readout if it's connected to TB3. When connected to TB1, M1539 did not show any problem.
M1606 worked properly both with TB1 and TB3.
For FT test the configuration is following:
Entry  Fri Aug 28 14:07:24 2020, Andrey Starodumov, PhQualification, M1539 
There is no new PH optimisation for this module?!
To be checked!
Entry  Wed Nov 20 16:53:50 2019, Andrey , Full test, M1533 and M1534 FullQualification 
FullQualification of two modules M1533 and M1534 has been done. Full time of the test is about 7hrs.
Test included: FT@-10C, IV@-10, 2Cycles from -10C up to +10C, FT@-10C, IV@-10C, Ft@1-C and IV@10C.
No issues observed. 
Entry  Thu Feb 6 00:45:25 2020, Dinko Ferencek, Cold box tests, Lost at least one Peltier in the coldbox TemperatureCycle.pdf
During an overnight (Feb. 4-5) FullQualification run with modules 1545, 1547, 1548, and 1549 the coldbox lost the ability to maintain the temperature at
-20 C. As can be seen from the attached temperature log, the problems already started during the temperature cycles where the last 2 cycles were noticeably
longer than the first 3 and finally during the second Fulltests at -20 C the temperature started rising and stabilized around -10 C. Based on these observations
Entry  Fri Mar 27 14:32:11 2020, Andrey Starodumov, HDI test, List of suspicious HDIs 
Here are HDIs that the first time tested showed 500-600V measured on the pin or directly on the HDI pad instead of 800V:
5006, 5007, 2005
3014, 3017, 3019, 6021
Entry  Mon Jan 20 21:33:20 2020, Dinko Ferencek, Cold box tests, Latest tests of PH optimization TrimBitTest_old.pngTrimBitTest_new.png
We did some tests by running the latest version of pXar on modules M1521, M1529, M1534 and M1536, specifically trimming and PH optimization. Of the 4 modules
tested, the PH optimization converged only for M1521. We plan to further investigate with Urs' help possible reasons for failed PH optimization.
    Reply  Tue Jan 21 12:19:08 2020, Dinko Ferencek, Cold box tests, Latest tests of PH optimization 
[quote="Dinko Ferencek"]We did some tests by running the latest version of pXar on modules M1521, M1529, M1534 and M1536, specifically trimming and PH optimization.
Of the 4 modules tested, the PH optimization converged only for M1521. We plan to further investigate with Urs' help possible reasons for failed PH optimization.
Entry  Wed May 13 17:57:45 2020, Andrey Starodumov, Other, L1_DATA backup 
L1_DATA files are backed up to the LaCie disk
Entry  Fri Jan 24 18:22:00 2020, Dinko Ferencek, General, L1 replacement web page updated 
This week several new links were added to the main L1 replacement web page located at the following URL
[URL=http://cms.web.psi.ch/L1Replacement/]http://cms.web.psi.ch/L1Replacement/[/URL]
Entry  Fri Oct 23 13:33:04 2020, danek kotlinski, Module transfer, L1 & L2 modules to go to P5 
The following modules will be transported to CERN/P5 as spares.

1) L1 
Entry  Thu Jan 23 14:54:45 2020, Dinko Ferencek, Cold box tests, Keithley exchanged 
Yesterday during an attempt to run the FullQualification, elComandante would lose control over Keithley after performing the IV curve measurements and would
go in the subsequent Fulltest with HV off. This happened both between the two sets of tests at -20 C and between the tests at -20 C and +10 C. Today we
exchanged the old Keithley with a new one and we observe that the communication with this new Keithley is much smoother without any error codes and warning
Entry  Tue Aug 6 14:30:00 2019, Dinko Ferencek, Documentation, Jumo Imago 500 manuals 
Manuals can be found at https://www.manualslib.com/products/Jumo-Imago-500-8786441.html
Entry  Fri Mar 27 10:54:04 2020, Urs Langenegger, Full test, Issues with pc11366 on March 27 
On March 27 I had a lot of troubles getting the full qualification up
and running.
Entry  Thu Oct 31 10:41:12 2019, Matej Roguljic, Cold box tests, Issue with DTB_WWVASW 
On 29th of October, an issue with DTB_WWVASW was noticed when trying to run FullQualification on 4 modules at the same time. On a tested, working module,
setting Vana works but taking tornado plots (setvthrcomp) doesn't as well as other tests. Curiously enough, after switching the DTB with the one used for
HDI testing (DTB_WRN13L) the problem was still there. This lead us to believe that the problem was with cables or ground. All cables and the adapter were