CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layer 1 Replacement Elog, Page 8 of 16  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Category Subjectdown Text Attachments
  259   Mon May 11 14:40:16 2020 danek kotlinskiOtherM1582On Friday I have tested the module M1582 at
room temperature in the blue box.
The report in MoreWeb says that this module
 m1582_roc1_thr_1d.pngm1582_roc1_thr_2d.pngm1582_roc1_ph70.png 
  92   Thu Feb 27 10:14:09 2020 danek kotlinskiModule transferM1562 to DESYM1562 was not qualified during a full test
since the test did not complete.
Looking at the module I see that roc 2 has
  
  298   Fri Aug 28 11:53:23 2020 Andrey StarodumovPhQualificationM1555There is no new PH optimisation for this module?!
To be checked!
  
  280   Fri May 29 15:41:40 2020 Andrey StarodumovGeneralM1546ROC10 of M1546 has 107 pixels with trimming
problems:
in a VCal threshold scan after trimming 81
 C10-p10.pdfTrimBits.pdfC10-50.pdf 
  161   Tue Mar 31 06:54:12 2020 danek kotlinskiModule gradingM1542I have tested M1542.
It looks not bad but there is a group of
pixels in the lower/left corner of ROC6 which
 Canvas_1.png 
  237   Wed Apr 29 08:48:06 2020 Urs LangeneggerOtherM1539M1539 showed no readout. I tried, all without
success, 
- reconnecting the cable to the adapter multiple
  
  257   Mon May 11 13:19:51 2020 Andrey StarodumovCold box testsM1539After several attempts including reconnecting
the cable, M1539 had no readout if it's connected
to TB3. When connected to TB1, M1539 did
  
  300   Fri Aug 28 14:07:24 2020 Andrey StarodumovPhQualificationM1539There is no new PH optimisation for this module?!
To be checked!
  
  34   Wed Nov 20 16:53:50 2019 Andrey Full testM1533 and M1534 FullQualificationFullQualification of two modules M1533 and
M1534 has been done. Full time of the test
is about 7hrs.
  
  67   Thu Feb 6 00:45:25 2020 Dinko FerencekCold box testsLost at least one Peltier in the coldboxDuring an overnight (Feb. 4-5) FullQualification
run with modules 1545, 1547, 1548, and 1549
the coldbox lost the ability to maintain
 TemperatureCycle.pdf 
  149   Fri Mar 27 14:32:11 2020 Andrey StarodumovHDI testList of suspicious HDIsHere are HDIs that the first time tested showed
500-600V measured on the pin or directly
on the HDI pad instead of 800V:
  
  47   Mon Jan 20 21:33:20 2020 Dinko FerencekCold box testsLatest tests of PH optimizationWe did some tests by running the latest version
of pXar on modules M1521, M1529, M1534 and
M1536, specifically trimming and PH optimization.
 TrimBitTest_old.pngTrimBitTest_new.png 
  48   Tue Jan 21 12:19:08 2020 Dinko FerencekCold box testsLatest tests of PH optimization[quote="Dinko Ferencek"]We did some tests
by running the latest version of pXar on
modules M1521, M1529, M1534 and M1536, specifically
  
  264   Wed May 13 17:57:45 2020 Andrey StarodumovOtherL1_DATA backupL1_DATA files are backed up to the LaCie disk   
  55   Fri Jan 24 18:22:00 2020 Dinko FerencekGeneralL1 replacement web page updatedThis week several new links were added to
the main L1 replacement web page located
at the following URL
  
  304   Fri Oct 23 13:33:04 2020 danek kotlinskiModule transferL1 & L2 modules to go to P5The following modules will be transported
to CERN/P5 as spares.
  
  51   Thu Jan 23 14:54:45 2020 Dinko FerencekCold box testsKeithley exchangedYesterday during an attempt to run the FullQualification,
elComandante would lose control over Keithley
after performing the IV curve measurements
  
  1   Tue Aug 6 14:30:00 2019 Dinko FerencekDocumentationJumo Imago 500 manualsManuals can be found at https://www.manualslib.com/products/Jumo-Imago-500-8786441.html   
  147   Fri Mar 27 10:54:04 2020 Urs LangeneggerFull testIssues with pc11366 on March 27On March 27 I had a lot of troubles getting
the full qualification up
and running.
  
  31   Thu Oct 31 10:41:12 2019 Matej RoguljicCold box testsIssue with DTB_WWVASWOn 29th of October, an issue with DTB_WWVASW
was noticed when trying to run FullQualification
on 4 modules at the same time. On a tested,
  