Mon Apr 6 17:18:02 2020, Andrey Starodumov, Full test, FT of M1606, M1630, M1655, M1566
|
M1606: B due to mean noise of several ROCs> 200e
M1639: C* due to failure many pixels of ROC1 at +10C as before: to be run at +10C with CtrlReg=17
M1655: B due to mean noise of several ROCs> 200e
|
Tue Apr 7 14:26:48 2020, Andrey Starodumov, Reception test, RT of M1567 failed
|
After exchange of TBM1 the results is the same:
[B]WAS[/B]:
|
Tue Apr 7 15:25:21 2020, Andrey Starodumov, Reception test, RT of M1662 failed
|
Address decoding of M1662 failed in one double column of ROC4.
To be decided what to do with this module.
Currently in C* tray. |
Tue Apr 7 16:57:01 2020, Andrey Starodumov, Full test, FT of M1593, M1658, M1659, M1660
|
M1593: B due to Rel.gain width and mean noise of a few ROCs
M1658: B due to threshold and mean noise of ROC15
M1659: B due to threshold and mean noise of a few ROCs
|
Tue Apr 7 17:59:27 2020, Andrey Starodumov, HDI test, 6 HDIs tested
|
HDIs
2030, 2031, 1021, 1022, 1023, 1035
are tested. All OK. |
Tue Apr 7 18:02:07 2020, Andrey Starodumov, Reception test, RT of 1654
|
After Silvan removed a cap and re-bonded broken and bent wires M1654 again has been tested.
RT grade is A.
Protection cap to be glued and to be (F)tested. |
Wed Apr 8 13:40:45 2020, Andrey Starodumov, Reception test, RT of M1625 failed
|
Silvan substituted TBM on this module.
Now ROC0 does not have hits.
Grade C, goes to "Bad" tray. |
Wed Apr 8 13:43:15 2020, Andrey Starodumov, Reception test, RT of M1623 failed
|
M1623: all ROCs are programmable but no readout from ROC8-ROC11.
Visual inspection is OK.
To module doctor. |
Wed Apr 8 15:02:37 2020, Andrey Starodumov, Software, Change of CrtlReg for RT
|
So far in a blue box for RT tbm10d_procv3 parameters were used.
This is wrong, since CtrlReg=9 is better for -20C while for +10 or higher T
CtrlReg=17 is better.
|
Wed Apr 8 15:20:38 2020, Andrey Starodumov, Reception test, RT of M1665
|
With CtrlReg=9 RT grade was B due to 90 noisy pixels in ROC5.
Noisy in this case means that one pixel in a 2x2 cluster in a few column got 40 hits instead of 10.
With CtrlReg=17 this problem gone. RT grade is A. |
Wed Apr 8 17:09:05 2020, Andrey Starodumov, Full test, FT of M1572, M1661, M1663, M1664
|
M1572: Grade B due to mean noise of ROC4 211 electrons
M1661: Grade B due to mean noise of a few ROCs > 200 electrons and in ROC12 44 pixels failed threshold cut
M1663: Grade B due to mean noise of a few ROCs > 200 electrons
|
Wed Apr 8 17:17:13 2020, Andrey Starodumov, Reception test, RT of M1662 failed
|
[quote="Andrey Starodumov"]Address decoding of M1662 failed in one double column of ROC4.
To be decided what to do with this module.
Currently in C* tray.[/quote]
|
Thu Apr 9 14:36:10 2020, Andrey Starodumov, Reception test, RT of M1669 and M1670
|
Both modules graded A. |
Thu Apr 9 14:49:07 2020, Andrey Starodumov, Reception test, RT of M1650 failed
|
Module has been tested on April 2nd.
Under Molex connector in ROC12 471 dead or noisy pixels.
Grade C. |
Thu Apr 9 17:24:49 2020, Andrey Starodumov, Full test, FT of M1654, M1665, M1666, M1667
|
M1654: Grade A
M1665: Grade B due to noisy pixels in ROC5. To be checked by module doctor!
M1666: Grade B due to mean noise of several chips > 200 electrons. Again on ROC12 there is a cluster of 31 dead bumps!
|
Thu Apr 9 17:36:31 2020, Andrey Starodumov, Reception test, RT of M1671 failed
|
ROC12-ROC15 no hits.
A candidate to TBM0 substitution, hence Grade C*.
To Module doctor1 |
Thu Apr 9 17:43:54 2020, Andrey Starodumov, HDI test, 4 HDIa tested
|
HDIs 3029, 3031, 3032, 4042 passed tests. All OK. |
Tue Apr 14 15:49:05 2020, Andrey Starodumov, Reception test, RT of M1623, M1657, M1673, M1674
|
M1623: Grade A, should be B due to 71 bump defects in ROC4
M1657: Grade B, due to 51 dead pixels in ROC12
M1673: Grade A, again 31 dead bumps in ROC12
|
Tue Apr 14 16:54:31 2020, Andrey Starodumov, Other, M1633 cable disconnected
|
We do not have any more module holders.
M1633 was in "Module doctor" tray.
I took module off the holder and put it in a gel-pak. |
Tue Apr 14 17:19:44 2020, Andrey Starodumov, Full test, FT of M1668, M1669, M1670, M1672
|
M1668: Grade B due to mean noise >200e for several ROCs
M1669: Grade B due to ROC2 mean noise >200e
M1670: Grade B due to ROC1 mean noise >200e
|