ID |
Date |
Author |
Category |
Subject |
Text |
|
139
|
Wed Mar 25 18:31:46 2020 |
Andrey Starodumov | Re-grading | Reanalised test results | [quote="Andrey Starodumov"]Test resulsts of
several modules have been re-analised without
grading on trimbit failure.
|
|
140
|
Wed Mar 25 18:40:53 2020 |
Andrey Starodumov | Full test | FT of M1599, M1613, M1624, M1616 | M1599: B due to leakage current at +10 (2-3umA)
M1613: B due to a few pixels with bad trimmed
threshold
|
|
144
|
Thu Mar 26 17:41:52 2020 |
Andrey Starodumov | Reception test | RT of M1629-1632 | All modules M1629, M1630, M1631 and M1632
graded A |
|
145
|
Thu Mar 26 17:43:06 2020 |
Andrey Starodumov | HDI test | 6+1 HDI tested | HDIs: 6021, 3014, 2005, 2009, 2006 and 2010
are OK
HDI 6024 failed: SDA0 and SDA2 on Ch1 missing
|
|
146
|
Thu Mar 26 17:45:15 2020 |
Andrey Starodumov | Full test | FT of M1545, M1557, M1627, M1628 | Retested M1545: C->B (to be correct on the
MoreWeb summary page)
Retested 1557: C->C in one ROC >160 pixels |
|
148
|
Fri Mar 27 14:28:14 2020 |
Andrey Starodumov | Reception test | M1633 failed Reception | All ROCs are programmable but permanent DESERALISER
ERROR, Ch6 and Ch7 event ID mismatch.
Visual inspection is OK
|
|
149
|
Fri Mar 27 14:32:11 2020 |
Andrey Starodumov | HDI test | List of suspicious HDIs | Here are HDIs that the first time tested showed
500-600V measured on the pin or directly
on the HDI pad instead of 800V:
|
|
150
|
Fri Mar 27 14:46:04 2020 |
Andrey Starodumov | Module doctor | Wolfram's summay from Mar 25 | M1542 nothing on sdata3 (12-15), bad output
or wire-bond
M1544 ROC15 timing off, can be made to |
|
151
|
Fri Mar 27 16:41:16 2020 |
Andrey Starodumov | Module doctor | Wolfram's summay from Mar 25 | [quote="Andrey Starodumov"]M1542 nothing
on sdata3 (12-15), bad output or wire-bond
M1544 ROC15 timing off, can be made to |
|
152
|
Fri Mar 27 17:27:09 2020 |
Andrey Starodumov | Reception test | M1635 failed Reception | All ROCs are programmable but DESER400 trailer
error bits: "NO DATA" or "IDLE DATA".
ROC8-11 are affected (no data)
|
|
153
|
Fri Mar 27 18:28:34 2020 |
Andrey Starodumov | Full test | 4 HDIs tested | HDIs 5020, 5018, 1044 and 1041 are OK |
|
154
|
Fri Mar 27 18:29:53 2020 |
Andrey Starodumov | Full test | FT of M1629, M1630, M1631, M1632 | M1629: B due to mean noise at -20C
M1630: C due to ROC1 with all pixels failed
of PH calibration (Gain). Both FT at -20C |
|
158
|
Mon Mar 30 16:46:59 2020 |
Andrey Starodumov | DB | Sensors missing in Advacam spreadsheet | Dinko noticed that several sensors are missing
in Advacam spreadsheet:
- 381785-03-3 used for M1586
|
|
159
|
Mon Mar 30 17:40:35 2020 |
Andrey Starodumov | HDI test | 3+2 HDIs (re-)tested | For 3 HDIs: 3014, 3017, 6021, HV has been
re-measured. In all cases 790V is measured
with 800V supplied.
|
|
160
|
Mon Mar 30 17:49:31 2020 |
Andrey Starodumov | Reception test | RT of M1637-1640 | All modules: M1637, M1638, M1639, M1649, are
graded A.
|
|
162
|
Tue Mar 31 08:21:39 2020 |
Andrey Starodumov | Full test | FT of 1558, 1606, 1634, 1636 on Mar 30 | M1558: B
M1606: C again due to trimmed threshold for
189 pixels of ROC2 at -20C. Second time at |
|
165
|
Tue Mar 31 17:32:47 2020 |
Andrey Starodumov | Reception test | RT of M1641-M1644 | All modules graded A. |
|
166
|
Tue Mar 31 17:33:37 2020 |
Andrey Starodumov | Reception test | RT of M1546 | This is a module with a broken TBM. Silvan
put a new one on top of the broken TBM.
The module is graded A after reception test.
|
|
168
|
Tue Mar 31 18:14:47 2020 |
Andrey Starodumov | HDI test | 3 HDI tested | HDIs: 1027, 5002, 5003 are tested. All OK. |
|
169
|
Tue Mar 31 18:23:18 2020 |
Andrey Starodumov | Full test | FT of M1637, M1638, M1639, M1640 | M1637: C. Graded C due to not completed first
test at -20C. Urs has reported issues. The