CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layer 1 Replacement Elog, Page 10 of 16  Not logged in ELOG logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Authordown Category Subject Text Attachments
  139   Wed Mar 25 18:31:46 2020 Andrey StarodumovRe-gradingReanalised test results[quote="Andrey Starodumov"]Test resulsts of
several modules have been re-analised without
grading on trimbit failure. 
  
  140   Wed Mar 25 18:40:53 2020 Andrey StarodumovFull testFT of M1599, M1613, M1624, M1616M1599: B due to leakage current at +10 (2-3umA)
M1613: B due to a few pixels with bad trimmed
threshold
  
  144   Thu Mar 26 17:41:52 2020 Andrey StarodumovReception testRT of M1629-1632All modules M1629, M1630, M1631 and M1632
graded A
  
  145   Thu Mar 26 17:43:06 2020 Andrey StarodumovHDI test6+1 HDI testedHDIs: 6021, 3014, 2005, 2009, 2006 and 2010
are OK
HDI 6024 failed: SDA0 and SDA2 on Ch1 missing
  
  146   Thu Mar 26 17:45:15 2020 Andrey StarodumovFull testFT of M1545, M1557, M1627, M1628Retested M1545: C->B (to be correct on the
MoreWeb summary page)
Retested 1557: C->C in one ROC >160 pixels
  
  148   Fri Mar 27 14:28:14 2020 Andrey StarodumovReception testM1633 failed ReceptionAll ROCs are programmable but permanent DESERALISER
ERROR, Ch6 and Ch7 event ID mismatch.
Visual inspection is OK
  
  149   Fri Mar 27 14:32:11 2020 Andrey StarodumovHDI testList of suspicious HDIsHere are HDIs that the first time tested showed
500-600V measured on the pin or directly
on the HDI pad instead of 800V:
  
  150   Fri Mar 27 14:46:04 2020 Andrey StarodumovModule doctorWolfram's summay from Mar 25M1542     nothing on sdata3 (12-15), bad output
or wire-bond
M1544     ROC15 timing off, can be made to
  
  151   Fri Mar 27 16:41:16 2020 Andrey StarodumovModule doctorWolfram's summay from Mar 25[quote="Andrey Starodumov"]M1542     nothing
on sdata3 (12-15), bad output or wire-bond
M1544     ROC15 timing off, can be made to
  
  152   Fri Mar 27 17:27:09 2020 Andrey StarodumovReception testM1635 failed ReceptionAll ROCs are programmable but DESER400 trailer
error bits: "NO DATA" or "IDLE DATA".
ROC8-11 are affected (no data)
  
  153   Fri Mar 27 18:28:34 2020 Andrey StarodumovFull test4 HDIs testedHDIs 5020, 5018, 1044 and 1041 are OK   
  154   Fri Mar 27 18:29:53 2020 Andrey StarodumovFull testFT of M1629, M1630, M1631, M1632M1629: B due to mean noise at -20C
M1630: C due to ROC1 with all pixels failed
of PH calibration (Gain). Both FT at -20C
  
  158   Mon Mar 30 16:46:59 2020 Andrey StarodumovDBSensors missing in Advacam spreadsheetDinko noticed that several sensors are missing
in Advacam spreadsheet:
- 381785-03-3 used for M1586
  
  159   Mon Mar 30 17:40:35 2020 Andrey StarodumovHDI test3+2 HDIs (re-)testedFor 3 HDIs: 3014, 3017, 6021, HV has been
re-measured. In all cases 790V is measured
with 800V supplied.
  
  160   Mon Mar 30 17:49:31 2020 Andrey StarodumovReception testRT of M1637-1640All modules: M1637, M1638, M1639, M1649, are
graded A.
 
  
  162   Tue Mar 31 08:21:39 2020 Andrey StarodumovFull testFT of 1558, 1606, 1634, 1636 on Mar 30M1558: B 
M1606: C again due to trimmed threshold for
189 pixels of ROC2 at -20C. Second time at
  
  165   Tue Mar 31 17:32:47 2020 Andrey StarodumovReception testRT of M1641-M1644All modules graded A.   
  166   Tue Mar 31 17:33:37 2020 Andrey StarodumovReception testRT of M1546This is a module with a broken TBM. Silvan
put a new one on top of the broken TBM.
The module is graded A after reception test.
  
  168   Tue Mar 31 18:14:47 2020 Andrey StarodumovHDI test3 HDI testedHDIs: 1027, 5002, 5003 are tested. All OK.   
  169   Tue Mar 31 18:23:18 2020 Andrey StarodumovFull testFT of M1637, M1638, M1639, M1640M1637: C. Graded C due to not completed first
test at -20C. Urs has reported issues. The
second -20C after T-Cycle and test at +10C
  
ELOG V3.1.3-7933898