  59   Mon Feb 3 15:35:09 2020 Anrey StarodumovFull test4 modules FT: 1539, 1541,1542, 1543
Jan 29
- Reception test for 1539 OK
- Cap gluing

Jan 30
- Reception test for 1541, 1542, 1543 OK
- Cap gluing
- FT:
-- 1539: C/B/B. First -20 is C due to INCOMPLETE TEST: DTB stuck
==> Must be retested at -20C and then files merging
-- 1541: C/C/C. Electrical Test is C due to Trimbit test failure in ROC13. Threshold is reasonable (see attachment)
==> Manually regraded to B
-- 1542: C/C/C. Electrical Test is C due to RelGain of 0.200 instead of <0.1 (too wide Gain distribution).
-- 1543: B/B/B (-20/-20/+10). Electrical Test is B due to mean noise >200e.
  58   Mon Feb 3 11:28:38 2020 Anrey StarodumovFull test4 modules FT: 1536, 1537,1538, 1540
Jan 28: Reception test for 1536, 1537, 1538 OK
Then cap gluing
Jan 29
- Reception test for 1540 OK
- FT:
-- 1536: B/B/B (-20/-20/+10). Electrical Test is B due to mean noise >200e. IV: 3umA for -20 and +10, slope >300
-- 1537: C/B/B. First -20 is C due ROC10 >200 TrimBits failure, noise >200e. IV: A
-- Trimbit test fails but Trimbit distribution is OK and Threshold reasonable (attachment)
==> Manually regraded to B
-- 1538: B/B/B (-20/-20/+10). Electrical Test is B due to mean noise >200e. IV: <0.2umA for +10, slope ~1
-- 1540: B/B/B (-20/-20/+10). Electrical Test is B due to mean noise >200e. IV: <0.8umA for +10, slope ~2.7
  43   Mon Dec 2 17:29:06 2019 Andrey StarodumovFull test4 modules FT
FullQualification of four modules M1526, M1529, M1521 and M1534 has been done. Full time of the test is about 5.5hrs.
Test included: FT@-20C, IV@-20, 3Cycles from -20C up to +10C, FT@10C and IV@10C.
Only M1521 passed the test (Grade B). Several issues observed in other modules:
1) 2-6 ROCs with TrimBit test failure for many pixels
2) Trimming id bad for some ROCs
3) test at +10C often better than at -20C
4) some ROCs hav issues with Thr and Gain distributions (out of specifications)

To be understood before the mass production!
  44   Tue Dec 3 12:15:54 2019 Andrey StarodumovFull test4 modules FT

Andrey Starodumov wrote:
FullQualification of four modules M1526, M1529, M1521 and M1534 has been done. Full time of the test is about 5.5hrs.
Test included: FT@-20C, IV@-20, 3Cycles from -20C up to +10C, FT@10C and IV@10C.
Only M1521 passed the test (Grade B). Several issues observed in other modules:
1) 2-6 ROCs with TrimBit test failure for many pixels
2) Trimming id bad for some ROCs
3) test at +10C often better than at -20C
4) some ROCs have issues with Thr and Gain distributions (out of specifications)

To be understood before the mass production!


We have run with Urs TrimBit test for M1526 at room T (without turning ON coldbox)
and results are confirmed yesterday observations of TrimBit test failure of several ROCs.
Urs will take a look and try to understand the issue.

To illustrate the problem below two summary tables of M1526 at -20C and +10C are shown.
  153   Fri Mar 27 18:28:34 2020 Andrey StarodumovFull test4 HDIs tested
HDIs 5020, 5018, 1044 and 1041 are OK
  221   Mon Apr 20 17:09:12 2020 Andrey StarodumovHDI test4 HDIs tested
After staying in 90%+ RH 2 HDIs became flat. The first one was easy to mount on an HDI holder.
But after 1-2hrs the second HDI became bent again, but still remained flexible, so was also easy to mount.
I put 2 more HDIs in the same conditions and after 2 hrs was able to mount and test them.
4041, 4043, 3033, 3041 tested. All OK.
  229   Thu Apr 23 18:01:16 2020 Andrey StarodumovHDI test4 HDIs tested
The following HDIs are tested:
6007: OK
1034: Failed due to not working Channel 1 in CLK0, CTR0, SDA0 and SDA1
6006: OK
5021: OK
  208   Thu Apr 9 17:43:54 2020 Andrey StarodumovHDI test4 HDIa tested
HDIs 3029, 3031, 3032, 4042 passed tests. All OK.
  159   Mon Mar 30 17:40:35 2020 Andrey StarodumovHDI test3+2 HDIs (re-)tested
For 3 HDIs: 3014, 3017, 6021, HV has been re-measured. In all cases 790V is measured with 800V supplied.
2 HDIs: 1026, 5022 were tested. Both OK.
All 5 HDIs will be used in module assembly.
  121   Fri Mar 20 18:09:47 2020 Andrey StarodumovHDI test3 HDIs tested
5008 failed, but most likely die to the pin head mis-alignment. Test to be repeated.
5007 and 5008: passed tests but during HV measurements (-800V) with a multymeter I heard high frequency noise and instead of -800V measure either -100V or -500V but the voltage jumped significantly. This effect is to be understood.
For the moment these HDIs will not be used in production.
  230   Fri Apr 24 13:56:11 2020 Andrey StarodumovHDI test3 HDIs tested
Following HDIs tested from the box "to be understood":
5021, 3019, 5008. All are fine.
  234   Mon Apr 27 14:21:17 2020 Andrey StarodumovHDI test3 HDIs tested
# remaining HDIs from "to be understood" box were tested after flattening them during weekend in RH=99.9% box.
6024, 4034 are OK
1039 bad: no data from A1 and A2 DTB outputs, flat output
  168   Tue Mar 31 18:14:47 2020 Andrey StarodumovHDI test3 HDI tested
HDIs: 1027, 5002, 5003 are tested. All OK.
  61   Mon Feb 3 17:09:36 2020 Andrey StarodumovReception test2 modules RT failed: 1544, 1546
Feb 03
- RT failed
-- 1544:
--- all ROCs are programmable
--- ROC14 no reliable Threshold found, ROC15 no threshold at all

-- 1546
--- all ROCs are programmable
--- no threshold found for all ROCs
  184   Fri Apr 3 18:06:11 2020 Andrey StarodumovHDI test2 HDIs tested
HDIs 6018 and 6019 are passed tests.
  284   Thu Jun 4 15:32:50 2020 Andrey StarodumovModule transfer18 Modules shipped to PSI
Quick check: Leakage current, set Vana, VthrCompCalDel and PixelAlive
T=+24C

Module Current@-150V Programmable Readout

M1569 -0.690uA OK OK
M1568 -0.530uA OK OK
M1566 -1.530uA OK OK
M1565 -1.430uA OK OK
M1554 -0.970uA OK OK
M1553 -1.300uA OK OK
M1552 -0.919uA OK OK
M1551 -1.310uA OK OK
M1550 -1.708uA OK OK
M1548 -0.470uA OK OK
M1547 -1.510uA OK OK
M1545 -0.770uA OK OK
M1543 -0.800uA OK OK
M1541 -0.750uA OK OK
M1540 -1.440uA OK OK
M1539 -0.680uA OK OK
M1537 -0.816uA OK OK
M1536 -4.270uA OK OK
  276   Thu May 28 14:38:43 2020 Andrey StarodumovModule transfer18 Modules shipped to ETHZ
1536, 1537, 1539, 1540, 1541,
1543, 1545, 1547, 1548, 1550,
1551, 1552, 1553, 1554, 1565,
1566, 1568, 1569
  262   Mon May 11 21:41:15 2020 Dinko FerencekSoftware17 to 10 C changes in the production overview page
0c513ab8: a few more updates on the main production overview page related to the 17 to 10 C change
  224   Wed Apr 22 17:28:42 2020 Andrey StarodumovHDI test11 HDIs tested
After keeping HDIs in a very high RH for a dew hours (24 is fine) became flat and could be fixed on HDI holder by vacuum.
11 HDIs were tested, all good:
3036, 3043, 3034, 3044, 1034, 6006, 6007, 6005,
6002, 6004, 6001
  23   Mon Oct 28 17:01:05 2019 Matej RoguljicSoftware Investigating the bug with the Keithley client
We took module 1529 and tried recreating the issue observed in the beginning of October. To do this in a reasonable amount of time, a "shorttest" procedure was defined which consists only of pretest and pixelalive. Three runs were taken

Run number 1: Shorttest@10,IV@10
Run number 2: Shorttest@10, IV@10,Cycle(n=1, between 10 and -10), Shorttest@10, IV@10
Run number 3: Shorttest@10, IV@10, Cycle(n=5, between 10 and -10), Shorttest@10, IV@10

In runs 1 and 2, IV was done from -5 to -155 in steps of 10
In run 3, IV was done from -5 to -405 in steps of 10

No issues were observed during the tests themselves.

Running MoReWeb shows the Temperature, Humidity and Sum of Currents plots while individual tests show only Pixel Alive map. IV plot is missing in the MoReWeb output, however, it is present in the ivCurve.log file. Tried investigating why IV is not shown, but couldn't get to the bottom of it.

Tomorrow we'll use M1529 and M1530 in a FullTest to check if the problem would appear.