Entry  Wed Mar 18 17:46:07 2020, Andrey Starodumov, Full test, FT of M1601-M1604 
M1601-M1604 passed full test
M1601: grade B
M1602-M1604: Grade C. The main reason is failed pixels during trimbit test. 
Entry  Thu Mar 19 17:17:27 2020, Andrey Starodumov, Reception test, M1609-M16012 
Reception test done and caps are glued to modules M1609-M1612.
M1611 graded B due to double column failure on ROC13. Others graded A.
Entry  Thu Mar 19 18:01:50 2020, Andrey Starodumov, Full test, FT of M1605-M1608 
Only M1607 graded B. All others are graded C due to trimbit test. 
M1606 should be looked carefully and may be retested since the threshold after trimming has strange features
(although not for all temperatures) that may mean that some trimbits really do not work.
Entry  Fri Mar 20 14:46:31 2020, Andrey Starodumov, Reception test, M1615 failed 
M1615 is programmable but "no working phases found"
Visual inspection of wire bonds - OK.
To module doctor!
Entry  Fri Mar 20 14:53:52 2020, Andrey Starodumov, Reception test, M1616 failed 
ROC10 of M1616 is not programmable, no readout from ROC10 and ROC11.
Visual inspection of wire bonds - OK
To module doctor!
Entry  Fri Mar 20 17:05:58 2020, Andrey Starodumov, Reception test, M1617 failed 
ROC8 of M1617 is programmable but no readout from it.
Silvan noticed that a corner of one ROC of this module is broken,
this is exactly ROC8.
Entry  Fri Mar 20 18:09:47 2020, Andrey Starodumov, HDI test, 3 HDIs tested 
5008 failed, but most likely die to the pin head mis-alignment. Test to be repeated.
5007 and 5008: passed tests but during HV measurements (-800V) with a multymeter I heard high frequency noise and instead of -800V measure either -100V
or -500V but the voltage jumped significantly. This effect is to be understood. 
Entry  Fri Mar 20 18:32:34 2020, Andrey Starodumov, Full test, FT of M1609-M1612 
M1609 and M1611 are graded C
M1610 and M1612 are graded B
Entry  Tue Mar 24 15:16:56 2020, Andrey Starodumov, HDI test, 8 HDIs tested on Mar 23 
1014, 1016, 2001, 2003, 
2004, 4013, 4015, 5014
All are good.
Entry  Tue Mar 24 15:18:59 2020, Andrey Starodumov, Reception test, M1621 failed Reception 
On M1621 ROC8 is not programmable.
Entry  Tue Mar 24 15:20:18 2020, Andrey Starodumov, Reception test, M1623 failed Reception 
M1623: all ROCs are programmable but no readout from ROC0-ROC3.
Visual inspection is OK.
To module doctor. 
Entry  Tue Mar 24 15:41:29 2020, Andrey Starodumov, Module grading, Change in MoreWeb GradingParameters.cfg  
On March 3d the Vcal calibration parameter has been changed:  
- StandardVcal2ElectronConversionFactor from 50 to 44 (electrons/VCal)
Entry  Tue Mar 24 16:08:58 2020, Andrey Starodumov, Reception test, M1625 failed Reception 
M1625: all ROCs are programmable but no readout from ROC0-ROC3.
The same symptom as for M1623.
Visual inspection is OK.
Entry  Tue Mar 24 18:09:07 2020, Andrey Starodumov, Full test, FT of M1615, M1619, M1620, M1622 
Test results have been analysed with modified code:
M1615: B
M1619: A
Entry  Tue Mar 24 18:11:52 2020, Andrey Starodumov, Re-grading, Reanalised test results 
Test resulsts of several modules have been re-analised without grading on trimbit failure. 
M1614: C->B
M1613: C->A
Entry  Wed Mar 25 14:11:35 2020, Andrey Starodumov, Reception test, RT of M1613 and M1614 on Mar 20 
Reception test for these modules have been done on Mar20.
Grading A for both modules.
Entry  Wed Mar 25 14:17:51 2020, Andrey Starodumov, Full test, FT of M1609, M1613, M1614, M1618 on Mar 23 
FT for these modules has been done on Mar 23
M1609: C
M1613: C
Entry  Wed Mar 25 14:44:37 2020, Andrey Starodumov, Reception test, RT of M1627 and M1628 
Both modules graded A.
    Reply  Wed Mar 25 17:03:11 2020, Andrey Starodumov, Full test, FT of M1615, M1619, M1620, M1622 
[quote="Andrey Starodumov"]Test results have been analysed with modified code:
M1615: B
M1619: A
Entry  Wed Mar 25 18:30:23 2020, Andrey Starodumov, HDI test, 4+2 HDI tested 
2 suspicious HDIs retested and found OK: 5006/5007
5013, 5016, 3017, 3019 are OK.