Entry  Thu Apr 30 15:16:58 2020, Andrey Starodumov, Full test, FT of M1540, M1541, M1543, M1547 
Modules tested om April 29
M1540: Grade B due to many (>1000) pixels failed trimming but only 70 are in "C-zone" for ROC0 at -20C [COLOR=red]-> retest!!![/COLOR]
M1541: Grade B due to mean noise >200e for a few ROCs
Entry  Mon May 4 14:18:20 2020, Andrey Starodumov, Full test, FT of M1540, 1549, 1571, 1598 
M1540: Grade A
M1549: Grade B due to mean noise >200e for ROC2 and 48 dead pixels in ROC5
M1571: Grade B due to mean noise >200e for many ROCs 
Entry  Tue May 12 13:29:27 2020, Andrey Starodumov, Full test, FT of M1539, M1582, M1606 
M1539: Grade B due to mean noise >200e in few ROCs
M1582: Grade B due to mean noise >200e in few ROCs and at -20C 137 pixels in ROC1 failed trimming. For P5 could use older test results (trim parameters)
of April 27 M20_1 when only 23 pixels in ROC1 failed trimming
Entry  Wed Apr 29 14:08:42 2020, Andrey Starodumov, Full test, FT of M1536, M1537, M1538 
M1536: Grade B due to mean noise >200e for ROC1
M1537: Grade B due to mean noise >200e for a few ROCs
M1538: Grade B due to mean noise >200e for a few ROCs and trimming failure for 70 pixels in ROC14 at -20C
Entry  Tue Mar 31 08:21:39 2020, Andrey Starodumov, Full test, FT of 1558, 1606, 1634, 1636 on Mar 30 
M1558: B 
M1606: C again due to trimmed threshold for 189 pixels of ROC2 at -20C. Second time at -20C and at +10C the number of failed pixels is 90+, hence grading
is B.
Entry  Sun Feb 9 19:34:55 2020, Dinko Ferencek, Full test, FT for 4 modules: 1557, 1558, 1559, 1560  

[B]1557:[/B] [COLOR=red]C/C/C[/COLOR] (-20/-20/+10). Mean noise >200e for some ROCs, dead trimbits, failed trimming and mean noise >300e in ROC 4. IV:
0.20 uA at +10 C, slope 1.08
Entry  Sat Feb 8 20:48:11 2020, Dinko Ferencek, Full test, FT for 4 modules: 1550, 1551, 1552, 1553  M1553_C11_AddressDecoding.pdfTemperatureCycle.pdf

[B]1550:[/B] [COLOR=darkblue]B/B/B[/COLOR] (-20/-20/+10). Electrical grade is B due to mean noise >200e for some ROCs. IV: 0.45 uA at +10 C, slope 1.09
[B]1551:[/B] [COLOR=red]C/C/C[/COLOR] (-20/-20/+10). Mean noise >200e for some ROCs, dead trimbits in ROC 13, but trimmed Vcal threshold distribution and
Entry  Fri Feb 7 20:35:44 2020, Dinko Ferencek, Full test, FT for 4 modules: 1545, 1547, 1548, 1549 TemperatureCycle.pdf

[B]1545:[/B] [COLOR=red]C/C/C[/COLOR] (-20/-20/+10). Problematic ROC 14: mean noise >320e, many dead trimbits, wide trimmed Vcal threshold distribution.
IV: 0.20 uA at +10 C, slope 1.09
Entry  Sun Feb 9 10:52:17 2020, Dinko Ferencek, Full test, FT for 4 modules: 1539, 1554, 1555, 1556 

[B]1539:[/B] [COLOR=darkblue]B/B/B[/COLOR] (-20/-20/+10). Electrical grade is B due to mean noise >200e for some ROCs. IV: 0.17 uA at +10 C, slope 1.08
[B]1554:[/B] [COLOR=darkblue]B/B/B[/COLOR] (-20/-20/+10). Electrical grade is B due to mean noise >200e for some ROCs. IV: 0.44 uA at +10 C, slope 1.80
Entry  Fri Feb 14 09:48:27 2020, Dinko Ferencek, Full test, FT for 12 modules: 1561-1576 (1563, 1567, 1572, 1575 excluded) 
[COLOR=blue]Feb. 11[/COLOR]

[B]1561:[/B] [COLOR=red]C/[/COLOR][COLOR=darkblue]B/B[/COLOR] (-20/-20/+10). Mean noise >200e for some ROCs, dead trimbits in ROC 6, but trimmed Vcal threshold
Entry  Fri Nov 29 17:06:26 2019, Dinko Ferencek, General, Dropbox folder on CERNBox set up 
A dropbox folder for elComandante tar files was set up in my CERNBox and synchronized with the following local folder on the lab PC

/media/disk1/DATA/L1_DATA_Backup/CERNBox_Dropbox/
Entry  Sun Jul 5 13:20:55 2020, Dinko Ferencek, General, Disk cleanup on the lab PC (pc11366.psi.ch) pxar_junk_home.txt
Home folder was cleaned up by deleting old pXar output .log and .root files. The full list is in the attached file. This released around 11 GB of disk space.

In addition, the following MoReWeb output folders which are no longer needed were deleted
Entry  Mon Jan 18 13:53:44 2021, Andrey Starodumov, Other, DTB tests 
M2217
--->DTB 154 (one of the red cold box setup):
- flat cable (with HV):
Entry  Thu Sep 26 22:09:14 2019, Dinko Ferencek, Software, DAC configuration update 
In accordance with the agreement made in an email thread initiated by Danek, the following changes to DAC settings for ROCs

vsh: 30 -> 8
Entry  Thu May 7 00:27:41 2020, Dinko Ferencek, Module grading, Comment about TrimBitDifference and its impact on the Trim Bit Test 
To expand on the following [URL=https://elrond.irb.hr/elog/Layer+1+Replacement/253]elog[/URL], on Mar. 24 Andrey changed the TrimBitDifference parameter
in Analyse/Configuration/GradingParameters.cfg from 2 to -2
Entry  Fri Jun 5 16:22:19 2020, Andrey Starodumov, General, Cleaning L1_DATA directory 
To be able to analyse Xray data and keep Total Production overview in a proper state we need to clean L1_DATA directory.
Reasons:
1) if one analyse Xray test results with flag -new, then all deleted with -d flag tests will be analysed again and Total production overview will have
Entry  Wed Apr 8 15:02:37 2020, Andrey Starodumov, Software, Change of CrtlReg for RT 
So far in a blue box for RT tbm10d_procv3 parameters were used. 
This is wrong, since CtrlReg=9 is better for -20C while for +10 or higher T
CtrlReg=17 is better. 
Entry  Wed Mar 18 15:16:47 2020, Andrey Starodumov, Software, Change in trimmig algorithm 
Urs yesterday modified the algorithm and tested it. From today we are using it. Main change that threshold for trimming is not any more fixed to VthrComp=79
but calculated based on the bottom tornado value + 20. CalDel is also optimised. This allows us to avoid failures in the trimbit tests due to too high
threshold, eg if the bottom of tornado close to 70 a fraction of pixels in a chip could have threshold above 70 and hence fail in the test. 
Entry  Mon May 25 17:24:05 2020, Andrey Starodumov, Software, Change in MoreWeb ColumnUniformityPerColumn.py 
In file
~/L1_SW/MoReWeb/Analyse/TestResultClasses/CMSPixel/QualificationGroup/XRayHRQualification$ emacs Chips/Chip/ColumnUniformityPerColumn/ColumnUniformityPerColumn.py
the high and low rates at which double column uniformity is checked are hard coded. Rates for L2 was there. Now correction is added:
Entry  Tue Mar 24 15:41:29 2020, Andrey Starodumov, Module grading, Change in MoreWeb GradingParameters.cfg  
On March 3d the Vcal calibration parameter has been changed:  
- StandardVcal2ElectronConversionFactor from 50 to 44 (electrons/VCal)