CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layer 1 Replacement Elog, Page 13 of 16  Not logged in ELOG logo
ID Date Authordown Category Subject
  212   Tue Apr 14 17:29:46 2020 Andrey StarodumovGeneralROC4 or ROC12 defects
Starting from M1650 almost every module has a cluster of dead pixels or broken bump bonds on
ROC4 or (more often) ROC12. The number of defects varied from 20 to 40.
It's almost excluded that such damage is made at PSI, since both of us: Silvan and me, first time connected the cable
to the module.
It would be interesting to check whether these bare modules arrived all together.
  215   Wed Apr 15 17:26:46 2020 Andrey StarodumovFull testFT of M1623, M1657, M1673, M1674
M1623: Grade B due to rel gain width, in ROC4 74 pixels failed trimming (Threshold) and mean noise >200e
M1657: Grade B due to 70 dead pixels in ROC12 and mean noise >200e
M1673: Grade B due to mean noise >200e in a few ROCs
M1674: Grade B due to mean noise >200e in a few ROCs
  217   Thu Apr 16 15:21:51 2020 Andrey StarodumovChange TBM Change TBMs on M1635, M1653, M1671
M1635: no data from ROC8-ROC11 => change TBM1
M1653: ROC12-ROC15 not programmable => change TBM0
M1671: no data from ROC12-ROC15 => change TBM0

Modules to be given to Silvan
  218   Thu Apr 16 17:31:16 2020 Andrey StarodumovFull testFT of M1662, M1675, M1676
M1662: Grade C due to failure of ROC4 in almost all tests: PixelAlive, PH calibration etc
Should be investigated and retested. At Reception PixelAlive etc was OK, only one double column showed problems
M1675: Grade B due to mean noise > 200e for several ROCs
M1676: Grade B due to mean noise > 200e for several ROCs
  219   Fri Apr 17 18:05:45 2020 Andrey StarodumovFull testFT of M1542, M1557, M1630, M1649 only at +10C
M1542 has grade C for relative gain width. Was tested with early versions of test SW with trim VCal=40 and not yet optimized PH optimization/calibration.
Other modules have grade C only at +10C. This time CtrlReg=17 instead of 9 is used/
M1542: Grade B due to 61 pixels failed Threshold criteria (trimming)
M1557: Grade B due to mean noise and NOT any more like in FT 214 pixels failed Threshold criteria
M1630: Grade B due to mean noise and NOT any more like in FT 3883 pixels in ROC1 failed Gain criteria
M1649: Grade B due to mean noise only in one ROC and NOT any more like in FT 270 pixels failed Threshold criteria
  220   Mon Apr 20 15:20:07 2020 Andrey StarodumovReception testChange TBMs on M1635, M1653, M1671

Andrey Starodumov wrote:
M1635: no data from ROC8-ROC11 => change TBM1
M1653: ROC12-ROC15 not programmable => change TBM0
M1671: no data from ROC12-ROC15 => change TBM0

Modules to be given to Silvan


After TBMs have been changed:
M1635: the same no data from ROC8-ROC11
M1653: reception test Grade A
M1671: the same no data from ROC12-ROC15

M1635 and M1671 to module doctor for final decision
  221   Mon Apr 20 17:09:12 2020 Andrey StarodumovHDI test4 HDIs tested
After staying in 90%+ RH 2 HDIs became flat. The first one was easy to mount on an HDI holder.
But after 1-2hrs the second HDI became bent again, but still remained flexible, so was also easy to mount.
I put 2 more HDIs in the same conditions and after 2 hrs was able to mount and test them.
4041, 4043, 3033, 3041 tested. All OK.
  222   Tue Apr 21 15:34:57 2020 Andrey StarodumovGeneralRetesting starts today
From today we will retest modules that have been tested with pXar SW versions earlier than March 18.
There were a few changes before this date:
1) trimming VCal: 40->50
2) threshold at which trim bit test is done
3) improvements in PH optimization algorithm

No changes in test algorithm have been introduced since March 18.

All modules have been tested with CtrlReg=9, for this several modules failed at +10C.
From now on for test CtrlReg=17 will be used.

Ft will be shorter: only one test at -20C, no T-cycling and one test at +10C. Leakage current will be measured up to 200V.
  223   Tue Apr 21 17:39:48 2020 Andrey StarodumovFull testFT of M1542, M1554, M1555, M1653
M1542: Grade C due to massive (>1000pixels in total) trimming failures at -20C in ROC11,14,15. There was no such problem at previous test when CtrlReg=9 was used, while for the present test CtrlReg=17 was used
M1554: Grade C due to massive (>1000pixels in total) trimming failures at -20C in ROC9,13. There was no such problem at previous test when CtrlReg=9 was used module was graded B, while for the present test CtrlReg=17 was used
M1555: Grade B due to 75 pixels had trimming failures at -20C in ROC10. There was no such problem at previous test when CtrlReg=9 was used, while for the present test CtrlReg=17 was used
M1653: Grade B due to >1% (~50) pixels had trimming failures at -20C in ROC5,12.
  224   Wed Apr 22 17:28:42 2020 Andrey StarodumovHDI test11 HDIs tested
After keeping HDIs in a very high RH for a dew hours (24 is fine) became flat and could be fixed on HDI holder by vacuum.
11 HDIs were tested, all good:
3036, 3043, 3034, 3044, 1034, 6006, 6007, 6005,
6002, 6004, 6001
  225   Wed Apr 22 17:41:11 2020 Andrey StarodumovFull testFT of M1542, M1554, M1555, M1653

Andrey Starodumov wrote:

M1542: Grade C due to massive (>1000pixels in total) trimming failures at -20C in ROC11,14,15. There was no such problem at previous test when CtrlReg=9 was used, while for the present test CtrlReg=17 was used
M1554: Grade C due to massive (>1000pixels in total) trimming failures at -20C in ROC9,13. There was no such problem at previous test when CtrlReg=9 was used module was graded B, while for the present test CtrlReg=17 was used
M1555: Grade B due to 75 pixels had trimming failures at -20C in ROC10. There was no such problem at previous test when CtrlReg=9 was used, while for the present test CtrlReg=17 was used
M1653: Grade B due to >1% (~50) pixels had trimming failures at -20C in ROC5,12.


Repeat test with CtrlReg=9. ONLY 2 TESTs: -20C, IV@-20C (upto 205V), +10C,IV@+10C (up to 205V)

Warning: at +10C the total leakage current of all modules = 9.7umA!? From yesterday the IV curves: each single module had the current less 0.5-1umA
M1542: Grade B due failure of 41 pixels in trimming of ROC6 at -20C. Grading at+10C is A!
M1554: Grade B due to mean noise >200e in ROC14 at both temperatures
M1555: Grade B due to mean noise >200e in 2 ROC at both temperatures
M1653: Grade B due failure of 45 pixels in trimming of ROC12 and mean noise >200e in ROC0 at both temperatures
  226   Thu Apr 23 13:34:52 2020 Andrey StarodumovFull testFT of M1542, M1554, M1555, M1653

Andrey Starodumov wrote:

Andrey Starodumov wrote:

M1542: Grade C due to massive (>1000pixels in total) trimming failures at -20C in ROC11,14,15. There was no such problem at previous test when CtrlReg=9 was used, while for the present test CtrlReg=17 was used
M1554: Grade C due to massive (>1000pixels in total) trimming failures at -20C in ROC9,13. There was no such problem at previous test when CtrlReg=9 was used module was graded B, while for the present test CtrlReg=17 was used
M1555: Grade B due to 75 pixels had trimming failures at -20C in ROC10. There was no such problem at previous test when CtrlReg=9 was used, while for the present test CtrlReg=17 was used
M1653: Grade B due to >1% (~50) pixels had trimming failures at -20C in ROC5,12.


Repeat test with CtrlReg=9. ONLY 2 TESTs: -20C, IV@-20C (upto 205V), +10C,IV@+10C (up to 205V)

Warning: at +10C the total leakage current of all modules = 9.7umA!? From yesterday the IV curves: each single module had the current less 0.5-1umA
M1542: Grade B due failure of 41 pixels in trimming of ROC6 at -20C. Grading at+10C is A!
M1554: Grade B due to mean noise >200e in ROC14 at both temperatures
M1555: Grade B due to mean noise >200e in 2 ROC at both temperatures
M1653: Grade B due failure of 45 pixels in trimming of ROC12 and mean noise >200e in ROC0 at both temperatures


According to IV curves currents at 150V are: 0.27umA, 0.44umA, 0.49umA, 0.16umA.
  227   Thu Apr 23 13:37:29 2020 Andrey StarodumovFull testFT of M1556, M1557, M1559, M1560
M1556: Grade B due to several ROCs mean noise >200e
M1557: Grade B due to several ROCs mean noise >200e and trimming failed for 60 pixel in ROC4 at -20C
M1559: Grade A
M1560: Grade B due to several ROCs mean noise >200e
  228   Thu Apr 23 17:26:51 2020 Andrey StarodumovFull testFT of M1561, M1564, M1565, M1566
M1561: Grade B due to several ROCs mean noise >200e
M1564: Grade B due to two ROCs mean noise >200e
M1565: Grade B due to two ROCs mean noise >200e
M1566: Grade B due to two ROCs mean noise >200e
  229   Thu Apr 23 18:01:16 2020 Andrey StarodumovHDI test4 HDIs tested
The following HDIs are tested:
6007: OK
1034: Failed due to not working Channel 1 in CLK0, CTR0, SDA0 and SDA1
6006: OK
5021: OK
  230   Fri Apr 24 13:56:11 2020 Andrey StarodumovHDI test3 HDIs tested
Following HDIs tested from the box "to be understood":
5021, 3019, 5008. All are fine.
  231   Fri Apr 24 13:59:43 2020 Andrey StarodumovFull testFT of M1568, M1569, M1570, M1571
M1568: Grade B due to mean noise >200e for a few ROCs and for ROC0 RelGainWidth(=0.1) is twice larger then for other ROCs
M1569: Grade B due to mean noise >200e for a few ROCs
M1570: Grade B due to mean noise >200e for a few ROCs
M1571: Grade C due to trimming for 190 pixels failed in ROC4 at +10C. This is not real failure, the first time this module has been tested the grade was B at 10C (while trimming was done for VCal=40)

M1571 goes to C* tray. Solution: either repeat the current test or test only at +10C and merge later.
  232   Mon Apr 27 13:23:47 2020 Andrey StarodumovFull testFT of M1573, M1574, M1576, M1577
Test has been done on April 24
M1573: Grade B due to mean noise >200e for a few ROCs
M1574: Grade B due to mean noise >200e for a few ROCs and trimming failed for >100 pixels and RelGainWidth too wide for ROC0
M1576: Grade B due to mean noise >200e for a few ROCs
M1577: Grade B due RelGainWidth too wide for ROC13 at +10C, at -20C graded A!
  233   Mon Apr 27 13:50:09 2020 Andrey StarodumovFull testFT of M1578, M1579, M1580, M1581
M1578: Grade B due to mean noise >200e for a few ROCs and 67 pixels failed trimming in ROC1 at -20C
M1579: Grade A
M1580: Grade B due to mean noise >200e for a few ROCs and 59/112 pixels failed trimming in ROC5/ROC8 at +10C
M1581: Grade B due to mean noise >200e and 120/120 pixels failed trimming in ROC8/ROC13 at both temperatures
  234   Mon Apr 27 14:21:17 2020 Andrey StarodumovHDI test3 HDIs tested
# remaining HDIs from "to be understood" box were tested after flattening them during weekend in RH=99.9% box.
6024, 4034 are OK
1039 bad: no data from A1 and A2 DTB outputs, flat output
ELOG V3.1.3-7933898