3
|
Tue Aug 6 16:01:17 2019 |
Matej Roguljic | Software | Wrong dac settings - elComandante | If it seems that elComandante is taking wrong
dac settings for tests like Reception test
or full qualification, one should remember |
|
4
|
Tue Aug 6 16:06:45 2019 |
Matej Roguljic | Other | Vsh and ctrlreg for v4 chips | The recommended default value for Vsh is 8,
but the current version of pXar has it as
30. One should remember this when making |
|
5
|
Wed Aug 7 11:49:07 2019 |
Matej Roguljic | Module assembly | HDI glue irradiation tests | Six different glues were used to glue the
cap on six dummy modules. They were all irradiated
to 120 MRad in Zagreb after which they were |
10x |
6
|
Wed Aug 7 12:24:14 2019 |
Matej Roguljic | Documentation | Activities 31.7.-9.8.2019. | 31.7.
Matej tested 4 HDIs (8030-8033 and 8010). |
|
7
|
Wed Aug 7 18:01:06 2019 |
Matej Roguljic | Cold box tests | PhOptimization problem | August 7 - we found out that PhOptimization
algorithm starts leaking memory if it fails
to find proper values. If running one setup, |
|
8
|
Tue Sep 10 15:18:20 2019 |
Matej Roguljic | Other | Modules 1504, 1505, 1520 irradiation report | Modules 1504, 1505 and 1520 were taken to
Zagreb for irradiation on 13.08.19. The goal
was to check the v4 behavior after irradiation. |
|
9
|
Thu Sep 12 10:32:27 2019 |
Matej Roguljic | General | Activities 10.9.-19.9.2019. | 10.9.
Modules 1504,1505,1520 brought to PSI after
irradiation to 1.2 MGy in Zagreb. All 3 were |
|
11
|
Wed Sep 18 15:45:45 2019 |
Matej Roguljic | Module assembly | HDI sparking under HV test | HDI [B]8010[/B] passed all test except the
HV. Under the HV test, some sparks could
be heard and even seen by eye on the right |
7x |
13
|
Thu Sep 19 14:20:07 2019 |
Matej Roguljic | General | Activities 10.9.-19.9.2019. | [quote="Matej Roguljic"]10.9.
Modules 1504,1505,1520 brought to PSI after
irradiation to 1.2 MGy in Zagreb. All 3 were |
|
15
|
Thu Sep 26 17:38:07 2019 |
Matej Roguljic | General | Activities 25.-26.9. | It was planned to test a batch of HDIs these
2 days but the shipment was late. Dinko and
Andrei were shown the HDI testing procedure. |
|
23
|
Mon Oct 28 17:01:05 2019 |
Matej Roguljic | Software | Investigating the bug with the Keithley client | We took module 1529 and tried recreating the
issue observed in the beginning of October.
To do this in a reasonable amount of time, |
|
25
|
Tue Oct 29 16:36:23 2019 |
Matej Roguljic | Module assembly | Assembly lab cap gluing jigs | There are two jigs for cap gluing in the assembly
lab, circled in red and blue on the photo
in the attachment. The one closer to the |
|
26
|
Tue Oct 29 16:38:32 2019 |
Matej Roguljic | Software | Investigating the bug with the Keithley client | [quote="Matej Roguljic"]We took module 1529
and tried recreating the issue observed in
the beginning of October. To do this in a |
|
27
|
Wed Oct 30 10:38:04 2019 |
Matej Roguljic | General | Activities 28.-31.10.2019. | 28.10. - Investigated a problem with the FullQualification
setup, described more in this [URL=https://elrond.irb.hr/elog/Layer+1+Replacement/23]note[/URL]
29.10. - Further investigated the aforementioned |
|
30
|
Wed Oct 30 16:54:58 2019 |
Matej Roguljic | Software | Investigating the bug with the Keithley client | [quote="Matej Roguljic"][quote="Matej Roguljic"]We
took module 1529 and tried recreating the
issue observed in the beginning of October. |
|
31
|
Thu Oct 31 10:41:12 2019 |
Matej Roguljic | Cold box tests | Issue with DTB_WWVASW | On 29th of October, an issue with DTB_WWVASW
was noticed when trying to run FullQualification
on 4 modules at the same time. On a tested, |
|
35
|
Wed Nov 27 08:39:24 2019 |
Matej Roguljic | Cold box tests | Bump bonding test investigation | The BB test on M1534 shows a lot of dead bunmps
on C10 and some on C1 and C0 as well. Putting
a source on top of the module shows that |
|
38
|
Thu Nov 28 15:58:20 2019 |
Matej Roguljic | General | Activities 26.-29.11.2019. | 26.11.
Prepared the tools for RD53A digital module |
|
46
|
Mon Jan 20 18:09:50 2020 |
Matej Roguljic | General | Activities 20.-31.01.2020. | 20.01.
Expanded the HDI testing setup to be able |
|
56
|
Mon Jan 27 17:50:48 2020 |
Matej Roguljic | Cold box tests | Failing modules at -20 | Friday, 24.1., it was noticed that the modules
are graded as 'C' in similar fashion. Module
is graded A/B at +10 degrees and then graded |
|