ID |
Date |
Author |
Category |
Subject |
Text |
|
270
|
Fri May 22 16:19:01 2020 |
Andrey Starodumov | Software | Change in MoreWeb GradingParameters.cfg | Xray noise:
grade B moved from 300e to 400e
grade C moved from 400e to 500e |
|
271
|
Fri May 22 17:09:37 2020 |
Andrey Starodumov | FTs for ETHZ | Module list1 | [COLOR=green]green[/COLOR]: correct in Total
production overview
black: to remove old entries/raws
|
|
272
|
Mon May 25 16:58:23 2020 |
Andrey Starodumov | XRay HR tests | a few commments | These is just to record the information:
1. measured hit rates at which Efficiency
and Xray hits Maps are taken are 40-50% of |
|
273
|
Mon May 25 17:24:05 2020 |
Andrey Starodumov | Software | Change in MoreWeb ColumnUniformityPerColumn.py | In file
~/L1_SW/MoReWeb/Analyse/TestResultClasses/CMSPixel/QualificationGroup/XRayHRQualification$
emacs Chips/Chip/ColumnUniformityPerColumn/ColumnUniformityPerColumn.py
|
|
276
|
Thu May 28 14:38:43 2020 |
Andrey Starodumov | Module transfer | 18 Modules shipped to ETHZ | 1536, 1537, 1539, 1540, 1541,
1543, 1545, 1547, 1548, 1550,
1551, 1552, 1553, 1554, 1565,
|
|
277
|
Thu May 28 14:40:57 2020 |
Andrey Starodumov | Module transfer | 8 modules shipped to PSI | Quick check: Leakage current, set Vana, VthrCompCalDel
and PixelAlive
Module Current@-150V Programmable Readout
|
|
278
|
Fri May 29 13:56:35 2020 |
Andrey Starodumov | Re-grading | M1630 | In ROC1 of M1630 gain calibration failed massively
(3883 pixels) at +10C with CtrlReg 9.
A special test of M1630 only at +10C and |
|
279
|
Fri May 29 15:04:04 2020 |
Andrey Starodumov | XRay HR tests | Analysis of HRT: M1555-M1561 and M1564 | Module HRtest VCal
calibration Grade
#defects max #noisy pix
|
|
280
|
Fri May 29 15:41:40 2020 |
Andrey Starodumov | General | M1546 | ROC10 of M1546 has 107 pixels with trimming
problems:
in a VCal threshold scan after trimming 81 |
|
281
|
Fri May 29 15:56:27 2020 |
Andrey Starodumov | FTs for ETHZ | Module list2 | [COLOR=green]green:[/COLOR] correct in Total
production overview
black: to remove old entries/rows
|
|
284
|
Thu Jun 4 15:32:50 2020 |
Andrey Starodumov | Module transfer | 18 Modules shipped to PSI | Quick check: Leakage current, set Vana, VthrCompCalDel
and PixelAlive
T=+24C
|
|
285
|
Fri Jun 5 13:47:11 2020 |
Andrey Starodumov | Module grading | M1613 | FT of this modules has been done twice: March
23 and March 25, in both cases with the final
test SW. On March 23 module was graded C |
|
286
|
Fri Jun 5 16:22:19 2020 |
Andrey Starodumov | General | Cleaning L1_DATA directory | To be able to analyse Xray data and keep Total
Production overview in a proper state we
need to clean L1_DATA directory.
|
|
288
|
Mon Jun 15 10:36:45 2020 |
Andrey Starodumov | FTs for ETHZ | Module list 3 | M1596 2020-04-28
M1597 2020-04-30
M1598 2020-05-04
|
|
296
|
Fri Aug 14 13:47:26 2020 |
Andrey Starodumov | General | new PH optimisation test and Total Overview | New PH optimisation is done at -20C as a FT,
hence the results of this test are added
to "DB" file as the second m20_1 test.
|
|
297
|
Fri Aug 28 11:13:33 2020 |
Andrey Starodumov | Module grading | M1615 | M1615: one pixel in ROC10 unmaskable hence
should be graded C. Otherwise the module
is of grade A
|
|
298
|
Fri Aug 28 11:53:23 2020 |
Andrey Starodumov | PhQualification | M1555 | There is no new PH optimisation for this module?!
To be checked! |
|
299
|
Fri Aug 28 12:02:48 2020 |
Andrey Starodumov | XRay HR tests | M1599 | ROC5 has eff=93.65% and should be graded C.
Somehow efficiency was not taken into account
for HR test grading??? |
|
300
|
Fri Aug 28 14:07:24 2020 |
Andrey Starodumov | PhQualification | M1539 | There is no new PH optimisation for this module?!
To be checked! |
|
303
|
Fri Sep 18 15:45:21 2020 |
Andrey Starodumov | Other | M2211 and M2122 | I made a mistake and instead M2122 used ID
M2211 in the .ini file.