  248   Mon May 4 14:18:20 2020 Andrey StarodumovFull testFT of M1540, 1549, 1571, 1598M1540: Grade A
M1549: Grade B due to mean noise >200e for
ROC2 and 48 dead pixels in ROC5
  
  247   Mon May 4 14:13:39 2020 Andrey StarodumovFull testFT of M1552, M1553, M1595, M1597FT on April 30th
M1552: Grade B due to mean noise >200e for
ROC7,8
  
  246   Fri May 1 19:34:01 2020 danek kotlinskiModule gradingM1582M1582 was classified as C because of 167 pixels
failing trimming in ROC1.
I have tested this module.
 m1582_roc1_thr.pngm1582_roc1_thr_2d.png 
  245   Thu Apr 30 17:33:04 2020 Andrey StarodumovSoftwareMoReWeb empty DAC plots[quote="Matej Roguljic"]Some of the DAC parameters
plots were empty in the total production
overview page. All the empty plots had the
  
  244   Thu Apr 30 17:24:57 2020 Dinko FerencekSoftwareMoReWeb empty DAC plots[quote="Matej Roguljic"]Some of the DAC parameters
plots were empty in the production overview
page. All the empty plots had the number
  
  243   Thu Apr 30 16:47:00 2020 Matej RoguljicSoftwareMoReWeb empty DAC plotsSome of the DAC parameters plots were empty
in the total production overview page. All
the empty plots had the number "35" in them
  
  242   Thu Apr 30 15:38:43 2020 danek kotlinskiModule transferM1635 & M1671 transferred to gel-packTwo bad modules have been placed in gel-packs:
1635 & 1671.
  
  241   Thu Apr 30 15:25:36 2020 Andrey StarodumovFull testFT of M1548, M1549, M1550, M1551M1548: Grade B due to mean noise >200e for
ROC11
M1549: Grade B due to mean noise >200e for
  
  240   Thu Apr 30 15:16:58 2020 Andrey StarodumovFull testFT of M1540, M1541, M1543, M1547Modules tested om April 29
M1540: Grade B due to many (>1000) pixels
failed trimming but only 70 are in "C-zone"
  
  239   Wed Apr 29 18:11:36 2020 Andrey StarodumovFull testFT of M1590, M1592, M1596, M1600Modules tested om April 28
M1590: Grade B  Grade B due to mean noise
>200e for a few ROCs
  
  238   Wed Apr 29 14:08:42 2020 Andrey StarodumovFull testFT of M1536, M1537, M1538M1536: Grade B due to mean noise >200e for
ROC1
M1537: Grade B due to mean noise >200e for
  
  237   Wed Apr 29 08:48:06 2020 Urs LangeneggerOtherM1539M1539 showed no readout. I tried, all without
success, 
- reconnecting the cable to the adapter multiple
  
  236   Tue Apr 28 18:11:45 2020 Andrey StarodumovFull testFT of M1586, M1587, M1588, M1589M1586: Grade B due to mean noise >200e for
a few ROCs
M1587: Grade B due to mean noise >200e for
  
  235   Tue Apr 28 18:06:41 2020 Andrey StarodumovFull testFT of M1582, M1583, M1584, M1585Modules tested om Apr 27
M1582: Grade C due to 167 pixels failed trimming
on ROC1 at +10C only. Previous test on Feb
  
  234   Mon Apr 27 14:21:17 2020 Andrey StarodumovHDI test3 HDIs tested# remaining HDIs from "to be understood" box
were tested after flattening them during
weekend in RH=99.9% box.
  
  233   Mon Apr 27 13:50:09 2020 Andrey StarodumovFull testFT of M1578, M1579, M1580, M1581M1578: Grade B due to mean noise >200e for
a few ROCs and 67 pixels failed trimming
in ROC1 at -20C
  
  232   Mon Apr 27 13:23:47 2020 Andrey StarodumovFull testFT of M1573, M1574, M1576, M1577Test has been done on April 24
M1573: Grade B due to mean noise >200e for
a few ROCs
  
  231   Fri Apr 24 13:59:43 2020 Andrey StarodumovFull testFT of M1568, M1569, M1570, M1571M1568: Grade B due to mean noise >200e for
a few ROCs and for ROC0 RelGainWidth(=0.1)
is twice larger then for other ROCs
  
  230   Fri Apr 24 13:56:11 2020 Andrey StarodumovHDI test3 HDIs testedFollowing HDIs tested from the box "to be
understood":
5021, 3019, 5008. All are fine.
  
  229   Thu Apr 23 18:01:16 2020 Andrey StarodumovHDI test4 HDIs testedThe following HDIs are tested:
6007: OK
1034: Failed due to not working Channel 1
  