ID |
Date |
Author |
Category |
Subject |
Text |
|
64
|
Tue Feb 4 19:52:22 2020 |
Dinko Ferencek | Cold box tests | Blue coldbox setup for reception tests commissioned | Today I continued where Matej left off with
commissioning the blue coldbox setup for
reception tests. All the necessary software |
|
65
|
Wed Feb 5 16:07:03 2020 |
Dinko Ferencek | Module assembly | Module assembly spreadsheet migrated to Google Sheets | Excel file stored on CERNBox containing the
module assembly spreadsheet has been replaced
with a Google Sheet [URL=https://docs.google.com/spreadsheets/d/12m2ESCLPH5AyWuuYV4KV1c5O_NEy7uP5PQKzVsdr3wQ/edit?usp=sharing]https://docs.google.com/spreadsheets/d/12m2ESCLPH5AyWuuYV4KV1c5O_NEy7uP5PQKzVsdr3wQ/edit?usp=sharing[/URL]. |
|
66
|
Wed Feb 5 16:18:30 2020 |
Dinko Ferencek | Module transfer | Module location tracking added to the assembly spreadsheet | To keep track of the module location, two
columns were added to the module assembly
spreadsheet [URL=https://docs.google.com/spreadsheets/d/12m2ESCLPH5AyWuuYV4KV1c5O_NEy7uP5PQKzVsdr3wQ/edit?usp=sharing]https://docs.google.com/spreadsheets/d/12m2ESCLPH5AyWuuYV4KV1c5O_NEy7uP5PQKzVsdr3wQ/edit?usp=sharing[/URL] |
|
67
|
Thu Feb 6 00:45:25 2020 |
Dinko Ferencek | Cold box tests | Lost at least one Peltier in the coldbox | During an overnight (Feb. 4-5) FullQualification
run with modules 1545, 1547, 1548, and 1549
the coldbox lost the ability to maintain |
|
68
|
Thu Feb 6 17:09:50 2020 |
Dinko Ferencek | Module assembly | Gluing stamp improvements | Silvan has made additional grooves on the
gluing stamp which now sits stably on top
of the module without any rocking motion |
|
69
|
Thu Feb 6 17:19:12 2020 |
Dinko Ferencek | Reception test | RT for 6 modules: 1551, 1552, 1553, 1554, 1555, 1556 | Today reception test was run for 6 modules
and looks OK for all module. The modules
were graded as follows:
|
|
70
|
Thu Feb 6 17:28:23 2020 |
Dinko Ferencek | Cold box tests | New Peltiers installed | Today new Peltiers were installed and tested.
Unfortunately, we had trouble reaching -20
C with this new set of Peltiers (they are |
|
71
|
Fri Feb 7 19:40:27 2020 |
Dinko Ferencek | Reception test | RT for 6 modules: 1557, 1558, 1559, 1560, 1561, 1562 | Reception test is done and all 6 modules were
graded [COLOR=darkgreen]A[/COLOR].
|
|
72
|
Fri Feb 7 20:35:44 2020 |
Dinko Ferencek | Full test | FT for 4 modules: 1545, 1547, 1548, 1549 |
[B]1545:[/B] [COLOR=red]C/C/C[/COLOR] (-20/-20/+10).
Problematic ROC 14: mean noise >320e, many |
|
73
|
Sat Feb 8 20:48:11 2020 |
Dinko Ferencek | Full test | FT for 4 modules: 1550, 1551, 1552, 1553 |
[B]1550:[/B] [COLOR=darkblue]B/B/B[/COLOR]
(-20/-20/+10). Electrical grade is B due |
|
74
|
Sun Feb 9 10:52:17 2020 |
Dinko Ferencek | Full test | FT for 4 modules: 1539, 1554, 1555, 1556 |
[B]1539:[/B] [COLOR=darkblue]B/B/B[/COLOR]
(-20/-20/+10). Electrical grade is B due |
|
75
|
Sun Feb 9 12:05:47 2020 |
Dinko Ferencek | Module assembly | Cap gluing | Today protective caps have been glued to modules
1561 and 1562. |
|
76
|
Sun Feb 9 18:36:13 2020 |
Dinko Ferencek | Module grading | Manual grading | The procedure for manual grading is described
in [URL=https://github.com/psi46/MoReWeb/pull/120]https://github.com/psi46/MoReWeb/pull/120[/URL].
|
|
77
|
Sun Feb 9 19:34:55 2020 |
Dinko Ferencek | Full test | FT for 4 modules: 1557, 1558, 1559, 1560 |
[B]1557:[/B] [COLOR=red]C/C/C[/COLOR] (-20/-20/+10).
Mean noise >200e for some ROCs, dead trimbits, |
|
78
|
Tue Feb 11 10:25:32 2020 |
Dinko Ferencek | Reception test | RT for 6 modules: 1563, 1564, 1565, 1566, 1567, 1568 | [COLOR=blue]Feb. 10[/COLOR]
[B]1563:[/B] Grade [COLOR=red]C[/COLOR], |
|
79
|
Tue Feb 11 17:56:06 2020 |
Dinko Ferencek | Cold box tests | Fulltest failure - psiAgente | During today's full qualification the 2nd
fulltest@-20 for TB1 (M1562) failed
|
|
80
|
Wed Feb 12 10:43:04 2020 |
Dinko Ferencek | Module transfer | 9 modules sent to ETH: 1539, 1545, 1547, 1548, 1549, 1550, 1551, 1552, 1553 | Yesterday the following modules were sent
to ETH:
1539, 1545, 1547, 1548, 1549, 1550, 1551, |
|
81
|
Wed Feb 12 11:15:04 2020 |
Dinko Ferencek | Module grading | Problem with dead trimbits understood | It was noticed that some modules are graded
[COLOR=red]C[/COLOR] because of typically
just one ROC having a large number of dead |
14x |
82
|
Wed Feb 12 15:40:13 2020 |
Dinko Ferencek | Reception test | RT for 3 modules: 1569, 1570, 1571; 1572 bad | [B]1569:[/B] Grade [COLOR=darkgreen]A[/COLOR]
[B]1570:[/B] Grade [COLOR=darkgreen]A[/COLOR]
[B]1571:[/B] Grade [COLOR=darkblue]B[/COLOR], |
|
83
|
Thu Feb 13 21:44:28 2020 |
Dinko Ferencek | Reception test | RT for 6 modules: 1573, 1574, 1575, 1576, 1577, 1578 | [B]1573:[/B] Grade [COLOR=darkgreen]A[/COLOR]
[B]1574:[/B] Grade [COLOR=darkblue]B[/COLOR],
I(150)/I(100) > 2 (4.63)
|
|