Entry  Tue Mar 31 18:14:47 2020, Andrey Starodumov, HDI test, 3 HDI tested 
HDIs: 1027, 5002, 5003 are tested. All OK.
Entry  Tue Mar 31 09:29:02 2020, Urs Langenegger, Full test, exchanged adapter for DTB_WXC03A 
I did not manage to get any r/o from the module connected to that adapter, also after exchanging the module.

So I exchanged the module adapter with the one from the blue box, and all issues with r/o were gone (with both modules) 
    Reply  Tue Mar 31 17:36:36 2020, Urs Langenegger, Full test, exchanged adapter for DTB_WXC03A 
[quote="Urs Langenegger"]I did not manage to get any r/o from the module connected to that adapter, also after exchanging the module.

So I exchanged the module adapter with the one from the blue box, and all issues with r/o were gone (with both modules) 
Entry  Tue Mar 31 17:32:47 2020, Andrey Starodumov, Reception test, RT of M1641-M1644 
All modules graded A.
Entry  Tue Mar 31 10:32:32 2020, Urs Langenegger, Full test, M1637 
Maybe M1637 has an issue: It got stuck twice at the same place with 

[09:53:16.043] <TB0>     INFO:    PixTestReadback::CalibrateIa()
Entry  Tue Mar 31 08:21:39 2020, Andrey Starodumov, Full test, FT of 1558, 1606, 1634, 1636 on Mar 30 
M1558: B 
M1606: C again due to trimmed threshold for 189 pixels of ROC2 at -20C. Second time at -20C and at +10C the number of failed pixels is 90+, hence grading
is B.
Entry  Tue Mar 31 06:54:12 2020, danek kotlinski, Module grading, M1542 Canvas_1.png
I have tested M1542.
It looks not bad but there is a group of pixels in the lower/left corner of ROC6 which behave different.
One can also see that something is not right in some PH optimisation plots.
Entry  Mon Mar 30 17:49:31 2020, Andrey Starodumov, Reception test, RT of M1637-1640 
All modules: M1637, M1638, M1639, M1649, are graded A.
 
Entry  Mon Mar 30 17:40:35 2020, Andrey Starodumov, HDI test, 3+2 HDIs (re-)tested 
For 3 HDIs: 3014, 3017, 6021, HV has been re-measured. In all cases 790V is measured with 800V supplied.
2 HDIs: 1026, 5022 were tested. Both  OK.
All 5 HDIs will be used in module assembly.
Entry  Mon Mar 30 16:46:59 2020, Andrey Starodumov, DB, Sensors missing in Advacam spreadsheet 
Dinko noticed that several sensors are missing in Advacam spreadsheet:
- 381785-03-3 used for M1586
- 381783-02-1 used for M1617
Entry  Fri Mar 27 18:29:53 2020, Andrey Starodumov, Full test, FT of M1629, M1630, M1631, M1632 
M1629: B due to mean noise at -20C
M1630: C due to ROC1 with all pixels failed of PH calibration (Gain). Both FT at -20C are A!
Should be understood and retested. Put in C* tray.
    Reply  Mon Mar 30 14:52:59 2020, Danek Kotlinski, Full test, FT of M1629, M1630, M1631, M1632 m1630_roc1_ph_fits.png
I have tested M1630.
I see not problem with ROC1. See the attached plot.
There is one dcol in ROC12 which shoes the "pattern" problem seen in a few other ROCs.
    Reply  Mon Mar 30 15:40:52 2020, Urs, Full test, FT of M1629, M1630, M1631, M1632 phval-curve_M1630_p10_C1.pdfphshot_vcal255.pdfpixelalive_C1.pdf
M1630 is interesting because (I am using my terminology in the following) for the test at T=+10C ROC1 fails the PH optimization test and by consequence
the gain/pedestal test is also failed. 
Entry  Sun Mar 29 18:14:59 2020, danek kotlinski, Other, Modules 1544 & 1563 in gelpack 
The 2 bad modules:
M1544
M1563 
Entry  Fri Mar 27 18:28:34 2020, Andrey Starodumov, Full test, 4 HDIs tested 
HDIs 5020, 5018, 1044 and 1041 are OK
Entry  Fri Mar 27 17:27:09 2020, Andrey Starodumov, Reception test, M1635 failed Reception 
All ROCs are programmable but DESER400 trailer error bits: "NO DATA" or "IDLE DATA".
ROC8-11 are affected (no data)
Visual inspection is OK
Entry  Fri Mar 27 14:46:04 2020, Andrey Starodumov, Module doctor, Wolfram's summay from Mar 25 
M1542     nothing on sdata3 (12-15), bad output or wire-bond
M1544     ROC15 timing off, can be made to work (roc 15 vdig 11, ..), but probably not usable for the detector
M1545     full readout, ok, upgraded to C*
    Reply  Fri Mar 27 16:41:16 2020, Andrey Starodumov, Module doctor, Wolfram's summay from Mar 25 
[quote="Andrey Starodumov"]M1542     nothing on sdata3 (12-15), bad output or wire-bond
M1544     ROC15 timing off, can be made to work (roc 15 vdig 11, ..), but probably not usable for the detector
M1545     full readout, ok, upgraded to C*
Entry  Fri Mar 27 14:32:11 2020, Andrey Starodumov, HDI test, List of suspicious HDIs 
Here are HDIs that the first time tested showed 500-600V measured on the pin or directly on the HDI pad instead of 800V:
5006, 5007, 2005
3014, 3017, 3019, 6021
Entry  Fri Mar 27 14:28:14 2020, Andrey Starodumov, Reception test, M1633 failed Reception 
All ROCs are programmable but permanent DESERALISER ERROR, Ch6 and Ch7 event ID mismatch.
Visual inspection is OK
To module doctor
Entry  Fri Mar 27 10:54:04 2020, Urs Langenegger, Full test, Issues with pc11366 on March 27 
On March 27 I had a lot of troubles getting the full qualification up
and running.
Entry  Thu Mar 26 17:43:06 2020, Andrey Starodumov, HDI test, 6+1 HDI tested 
HDIs: 6021, 3014, 2005, 2009, 2006 and 2010 are OK
HDI 6024 failed: SDA0 and SDA2 on Ch1 missing
Entry  Thu Mar 26 17:41:52 2020, Andrey Starodumov, Reception test, RT of M1629-1632 
All modules M1629, M1630, M1631 and M1632 graded A
Entry  Thu Mar 26 10:41:24 2020, danek kotlinski, Module grading, M1606 
I have looked more closely at M1606.
Trimming show problems in ROC2, 8 pixels have 0 threshold and for 61 pixels thr=-1.
Strange because from Scurves the 61 have a high ~62 threshold but no failure.
    Reply  Thu Mar 26 15:37:18 2020, Urs, Module grading, M1606 anaTrim-thrDiff-data__M1606__pxar.pdf
attached find the threshold difference distributions for all four trimbits for all ROCs