  108   Mon Mar 16 10:05:23 2020 Matej RoguljicSoftwarePhQualification changeUrs made a change in pXar, in the PhOptimization
algorithm. One of the changes is in the testParameters.dat
where vcalhigh is set to 100 instead of 255.
  
  107   Sat Mar 14 17:08:14 2020 Matej RoguljicFull testFulltests on 2020/03/13
Modules tested:
M1591 C (gain at -20)
  
  105   Thu Mar 12 17:21:59 2020 Matej RoguljicFull test Fulltests on 2020/03/12 Modules tested:
M1557 C (trim bits at -20)
M1558 C (gain at -20)
  
  104   Thu Mar 12 12:19:43 2020 Matej RoguljicHDI test5 HDI tests on 12.3.I tested 5 HDIs on 12.3. 1030,1029,1040,1039
and 1038. 1039 failed all the electrical
tests. All the others passed the tests. HDI
  
  103   Wed Mar 11 17:59:48 2020 Andrey StarodumovRe-gradingRegrading C M1542: due to NoiseFollow the instruction from M1591 regrading
log:
"To correct the Summary page one needs to
  
  102   Wed Mar 11 17:22:57 2020 Matej RoguljicHDI test8 HDI tests on 11.3.I tested 8 HDIs today: 5029,5030,5031,6034,6036,6035,1031
and 1032
All of them tested fine.
  
  101   Wed Mar 11 17:12:05 2020 Matej RoguljicHDI testHDI testing procedure changeThere is an additional test that will be used
from 11.03. on all HDIs. It involves measuring
the voltage between ground and the HV pin
  
  100   Wed Mar 11 16:48:10 2020 Andrey StarodumovOtherM1586: issues with MOLEX?[quote="Urs Langenegger"]Module M1586 had
passed the full qualification on 20/02/27.
I had had to re-insert the cable in the Molex
  
  99   Wed Mar 11 15:34:57 2020 Urs LangeneggerOtherM1586: issues with MOLEX?Module M1586 had passed the full qualification
on 20/02/27. I had had to re-insert the cable
in the Molex connector for it to become programmable.
  
  98   Mon Mar 9 16:16:48 2020 Andrey StarodumovHDI testHDIs: 3001, 3002, 3003, 3004 and 4033Today I tested 5 HDIs: 3001, 3002, 3003, 3004
and 4033
Results:
  
  97   Wed Mar 4 16:00:05 2020 Andrey StarodumovRe-gradingRegrading C M1591: due to Noise[quote="Andrey Starodumov"]It has been realized
by Urs that VCal to electron conversion used
by MoreWeb is still 50e/VCal.
  
  96   Wed Mar 4 11:32:42 2020 danek kotlinskiFull testchange the target trim threshodl to vcal=50After some discussion we decided to change
the target trimming threshold from vcal 40
to 50.
  
  95   Tue Mar 3 14:05:13 2020 Andrey StarodumovRe-gradingRegrading C modules: due to NoiseIt has been realized by Urs that VCal to electron
conversion used by MoreWeb is still 50e/VCal.
While recent calibration done by Maren with
  
  94   Mon Mar 2 17:39:36 2020 Urs LangeneggerFull testFulltests on 2020/03/02Modules tested: 
M1595 C
M1596 B
  
  93   Fri Feb 28 17:33:16 2020 Urs LangeneggerFull testFulltests on 2020/02/28M1589 B
M1590 C (gain issues?)
M1591 C (noise issues?)
  
  92   Thu Feb 27 10:14:09 2020 danek kotlinskiModule transferM1562 to DESYM1562 was not qualified during a full test
since the test did not complete.
Looking at the module I see that roc 2 has
  
  91   Thu Feb 27 09:47:51 2020 Urs LangeneggerFull testFulltest on 2020/02/26On Wednesday, 2020/02/26 the following modules
went through the full qualification procedure:
  
  90   Wed Feb 26 10:34:55 2020 Urs LangeneggerFull testFulltest after Peltier replacementOn 2020/02/25, after SIlvan had replaced the
Peltiers, I did a fulltest. Here the summary:
  
  89   Tue Feb 25 17:11:18 2020 Urs LangeneggerModule assemblyModules glued and tested Feb 24/25Modules glued and tested Feb 24/25

The letters after the module name indicate
  
  88   Mon Feb 24 14:01:29 2020 Urs LangeneggerModule assemblyCap gluing in W8In week 8 the following modules were glued
and tested: 
  