CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layer 1 Replacement Elog, Page 8 of 16  Not logged in ELOG logo
ID Date Author Category Subject Text Attachments
  168   Tue Mar 31 18:14:47 2020 Andrey StarodumovHDI test3 HDI testedHDIs: 1027, 5002, 5003 are tested. All OK.   
  167   Tue Mar 31 17:36:36 2020 Urs LangeneggerFull testexchanged adapter for DTB_WXC03A[quote="Urs Langenegger"]I did not manage
to get any r/o from the module connected
to that adapter, also after exchanging the
  
  166   Tue Mar 31 17:33:37 2020 Andrey StarodumovReception testRT of M1546This is a module with a broken TBM. Silvan
put a new one on top of the broken TBM.
The module is graded A after reception test.
  
  165   Tue Mar 31 17:32:47 2020 Andrey StarodumovReception testRT of M1641-M1644All modules graded A.   
  164   Tue Mar 31 10:32:32 2020 Urs LangeneggerFull testM1637Maybe M1637 has an issue: It got stuck twice
at the same place with 
  
  163   Tue Mar 31 09:29:02 2020 Urs LangeneggerFull testexchanged adapter for DTB_WXC03AI did not manage to get any r/o from the module
connected to that adapter, also after exchanging
the module.
  
  162   Tue Mar 31 08:21:39 2020 Andrey StarodumovFull testFT of 1558, 1606, 1634, 1636 on Mar 30M1558: B 
M1606: C again due to trimmed threshold for
189 pixels of ROC2 at -20C. Second time at
  
  161   Tue Mar 31 06:54:12 2020 danek kotlinskiModule gradingM1542I have tested M1542.
It looks not bad but there is a group of
pixels in the lower/left corner of ROC6 which
 Canvas_1.png 
  160   Mon Mar 30 17:49:31 2020 Andrey StarodumovReception testRT of M1637-1640All modules: M1637, M1638, M1639, M1649, are
graded A.
 
  
  159   Mon Mar 30 17:40:35 2020 Andrey StarodumovHDI test3+2 HDIs (re-)testedFor 3 HDIs: 3014, 3017, 6021, HV has been
re-measured. In all cases 790V is measured
with 800V supplied.
  
  158   Mon Mar 30 16:46:59 2020 Andrey StarodumovDBSensors missing in Advacam spreadsheetDinko noticed that several sensors are missing
in Advacam spreadsheet:
- 381785-03-3 used for M1586
  
  157   Mon Mar 30 15:40:52 2020 UrsFull testFT of M1629, M1630, M1631, M1632M1630 is interesting because (I am using my
terminology in the following) for the test
at T=+10C ROC1 fails the PH optimization
 phval-curve_M1630_p10_C1.pdfphshot_vcal255.pdfpixelalive_C1.pdf 
  156   Mon Mar 30 14:52:59 2020 Danek KotlinskiFull testFT of M1629, M1630, M1631, M1632I have tested M1630.
I see not problem with ROC1. See the attached
plot.
 m1630_roc1_ph_fits.png 
  155   Sun Mar 29 18:14:59 2020 danek kotlinskiOtherModules 1544 & 1563 in gelpackThe 2 bad modules:
M1544
M1563 
  
  154   Fri Mar 27 18:29:53 2020 Andrey StarodumovFull testFT of M1629, M1630, M1631, M1632M1629: B due to mean noise at -20C
M1630: C due to ROC1 with all pixels failed
of PH calibration (Gain). Both FT at -20C
  
  153   Fri Mar 27 18:28:34 2020 Andrey StarodumovFull test4 HDIs testedHDIs 5020, 5018, 1044 and 1041 are OK   
  152   Fri Mar 27 17:27:09 2020 Andrey StarodumovReception testM1635 failed ReceptionAll ROCs are programmable but DESER400 trailer
error bits: "NO DATA" or "IDLE DATA".
ROC8-11 are affected (no data)
  
  151   Fri Mar 27 16:41:16 2020 Andrey StarodumovModule doctorWolfram's summay from Mar 25[quote="Andrey Starodumov"]M1542     nothing
on sdata3 (12-15), bad output or wire-bond
M1544     ROC15 timing off, can be made to
  
  150   Fri Mar 27 14:46:04 2020 Andrey StarodumovModule doctorWolfram's summay from Mar 25M1542     nothing on sdata3 (12-15), bad output
or wire-bond
M1544     ROC15 timing off, can be made to
  
  149   Fri Mar 27 14:32:11 2020 Andrey StarodumovHDI testList of suspicious HDIsHere are HDIs that the first time tested showed
500-600V measured on the pin or directly
on the HDI pad instead of 800V:
  
ELOG V3.1.3-7933898