CMS Pixel Detector Miscellaneous
Phase 1 Phase 2
Layer 1 Replacement Layers 2-4
  Layer 1 Replacement Elog, Page 2 of 16  Not logged in ELOG logo
Entry  Wed Jul 29 17:19:43 2020, danek kotlinski, PhQualification, Change configuration for PH qialification 
Preparing the new PH optimization I had to make the following modifications:

1) in elCommandante.ini
Entry  Fri Oct 23 13:33:04 2020, danek kotlinski, Module transfer, L1 & L2 modules to go to P5 
The following modules will be transported to CERN/P5 as spares.

1) L1 
    Reply  Fri Nov 6 07:28:42 2020, danek kotlinski, POS, POS configuration files created 
[quote="Dinko Ferencek"]pXar parameter files were converted to POS configuration files by executing the following commands on the lab PC at PSI

cd /home/l_tester/L1_SW/MoReWeb/scripts/
Entry  Mon Feb 24 13:53:32 2020, Urs Langenegger, Module assembly, Setup changes in week 8 
Setup changes in week 8

Silvan considered the single-module gluing approach suboptimal. It was changed: 
Entry  Mon Feb 24 14:01:29 2020, Urs Langenegger, Module assembly, Cap gluing in W8 
In week 8 the following modules were glued and tested: 

M1581
Entry  Tue Feb 25 17:11:18 2020, Urs Langenegger, Module assembly, Modules glued and tested Feb 24/25 
Modules glued and tested Feb 24/25

The letters after the module name indicate the reception test grade. 
Entry  Wed Feb 26 10:34:55 2020, Urs Langenegger, Full test, Fulltest after Peltier replacement 
On 2020/02/25, after SIlvan had replaced the Peltiers, I did a fulltest. Here the summary: 

  Finished all tests. Summary of test durations:
Entry  Thu Feb 27 09:47:51 2020, Urs Langenegger, Full test, Fulltest on 2020/02/26 
On Wednesday, 2020/02/26 the following modules went through the full qualification procedure: 
M1581
M1582
Entry  Fri Feb 28 17:33:16 2020, Urs Langenegger, Full test, Fulltests on 2020/02/28 
M1589 B
M1590 C (gain issues?)
M1591 C (noise issues?)
Entry  Mon Mar 2 17:39:36 2020, Urs Langenegger, Full test, Fulltests on 2020/03/02 
Modules tested: 
M1595 C
M1596 B
Entry  Wed Mar 11 15:34:57 2020, Urs Langenegger, Other, M1586: issues with MOLEX? 
Module M1586 had passed the full qualification on 20/02/27. I had had to re-insert the cable in the Molex connector for it to become programmable. 

On 2020/03/09, I tried to re-test M1586, but it was not programmable. Visual inspection revealed nothing to me. I did re-insert the cable once again, but
Entry  Fri Mar 27 10:54:04 2020, Urs Langenegger, Full test, Issues with pc11366 on March 27 
On March 27 I had a lot of troubles getting the full qualification up
and running.
Entry  Tue Mar 31 09:29:02 2020, Urs Langenegger, Full test, exchanged adapter for DTB_WXC03A 
I did not manage to get any r/o from the module connected to that adapter, also after exchanging the module.

So I exchanged the module adapter with the one from the blue box, and all issues with r/o were gone (with both modules) 
Entry  Tue Mar 31 10:32:32 2020, Urs Langenegger, Full test, M1637 
Maybe M1637 has an issue: It got stuck twice at the same place with 

[09:53:16.043] <TB0>     INFO:    PixTestReadback::CalibrateIa()
    Reply  Tue Mar 31 17:36:36 2020, Urs Langenegger, Full test, exchanged adapter for DTB_WXC03A 
[quote="Urs Langenegger"]I did not manage to get any r/o from the module connected to that adapter, also after exchanging the module.

So I exchanged the module adapter with the one from the blue box, and all issues with r/o were gone (with both modules) 
Entry  Wed Apr 29 08:48:06 2020, Urs Langenegger, Other, M1539 
M1539 showed no readout. I tried, all without success, 
- reconnecting the cable to the adapter multiple times
- connecting to the adapter in the blue box
Entry  Fri Jul 10 11:24:37 2020, Urs Langenegger, Reception test, proc600V3 modules 
This week I tested 12 modules built with proc600v3. The module numbers are M1722 - M1733. 
The results are summarized at the usual place: 
http://cms.web.psi.ch/L1Replacement/WebOutput/MoReWeb/Overview/Overview.html
    Reply  Thu Mar 26 15:37:18 2020, Urs, Module grading, M1606 anaTrim-thrDiff-data__M1606__pxar.pdf
attached find the threshold difference distributions for all four trimbits for all ROCs
    Reply  Mon Mar 30 15:40:52 2020, Urs, Full test, FT of M1629, M1630, M1631, M1632 phval-curve_M1630_p10_C1.pdfphshot_vcal255.pdfpixelalive_C1.pdf
M1630 is interesting because (I am using my terminology in the following) for the test at T=+10C ROC1 fails the PH optimization test and by consequence
the gain/pedestal test is also failed. 
Entry  Tue Aug 6 15:15:00 2019, Matej Roguljic, General, Activities 19.6.-4.7.2019. 
17.6.-21.6.

Andrei and Matej were at PSI working on the cap gluing setup
ELOG V3.1.3-7933898